LIFEPAK CR Plus/EXPRESS Defibrillator
Preventive Maintenance
Defibrillator Cleaning
WARNING!
Shock or fire hazard.
Do not immerse any portion of this defibrillator in
water or other fluids. Avoid spilling any fluids on the defibrillator or
accessories.
The defibrillator should be cleaned after each use to ensure proper device
operation before the next patient use.
Do not clean any part of the defibrillator or accessories with bleach, bleach
dilution, or phenolic compounds. Do not use abrasive or flammable cleaning
agents. Do not steam, autoclave, or gas-sterilize the defibrillator or its
accessories. Follow the cleaning guidelines provided below.
Item
Cleaning Method
Cleaning Agent
Defibrillator exterior case,
readiness display, and
crevices
Clean with damp
sponge or cloth
■
Nonabrasive soap and water
■
Quaternary ammonium compounds
■
Rubbing (isopropyl) alcohol
■
Peroxide (peracetic acid) solutions
CHARGE-PAK battery
charger
None
None — dispose of or recycle after use
(Continued on next page)
7-11