LIFEPAK CR Plus/EXPRESS Defibrillator
Preventive Maintenance
Maintenance and Inspection Schedule
Introduction
The defibrillator, as described within this service manual, is supported by a
warranty. For malfunctions encountered during use or inspection, contact
Physio-Control. In the USA, contact Physio-Control Technical Services at
1.800.442.1142 and select option 5 for service. Outside the USA, contact your
local Physio-Control representative.
The defibrillator does not require routine maintenance. The defibrillator performs
various automatic self-tests to ensure readiness. The results of these tests are
displayed in the
■
If the defibrillator passes the test, the
OK
indicator appears, indicating that
the device is ready for use.
■
If the defibrillator fails a test, the
BATTERY
indicator appears, indicating that
CHARGE-PAK battery charger needs to be replaced
.
■
If the defibrillator fails a test because it detects low device power, the
ATTENTION
indicator appears, indicating that the device has a low internal
battery. It may take from one to ten days before the battery is adequately
charged and the
OK
indicator appears.
■
If the defibrillator fails a test because it detects a fault, the
WRENCH
indicator
appears, indicating that the device should be removed from service.
(Continued on next page)