LIFEPAK CR Plus/EXPRESS Defibrillator
Modes of Operation
Auto Test Modes
The defibrillator periodically performs routine tests to detect problems that could
affect defibrillator performance. These tests include:
–
Initialization test
Initialization Test
The initialization test is performed each time the defibrillator is turned on (all the
defibrillator indicators will light momentarily, indicating that the test is in process).
The initialization test includes the following memory tests:
–
RAM test
–
Processor test
–
Watchdog timer test
If the initialization test completes successfully and verifies that it is not part of
another test, the defibrillator enters normal mode.
(Continued on next page)
5-3