Pulse output parameters
Current pulse shape
Biphase symmetrical (net current
0dc)
Operating conditions
Temperature
from 5°C to +40°C
Relative humidity
from 15% to 93% (non-condensing)
Atmospheric pressure
from 700hPa to 1060hPa
Storage and transport conditions
Temperature - electrodes
from 0°C to +40°C
Temperature - device
from -10°C to 50°C
Relative humidity
less than 93% (non-condensing)
Atmospheric pressure
from 700hPa to 1060hPa
Note: If you want to store the electrodes for more than
a month, keep them at temperatures b5°C
and +27°C.
Note: Do not store the electrodes in the freezer or
refrigerator. Do not leave them in extreme heat, nor
immerse them in water, or leave them outside of the
plastic bag.