Monitor Maintenance
Maintenance 13-7
M
a
int
e
na
nc
e
Preventative Maintenance
Replace backlight tube
assembly
Cleaning procedures
2.5 years of continuous use
or as needed
Perform the Preventive
Maintenance Procedure for
the M1180A #A01/A02 Wall
Mount Assembly every year.
See “Testing and
Maintaining the CMS” in the
“CMS Service Guide” or
“Testing and Maintaining
the V24 and V26“in the “V24
and V26 Service Guide”.
Full performance checks
including safety checks.
At least every two years,
after any repairs, or as
needed.
See “Testing and
Maintaining the CMS” in the
“CMS Service Guide” or
“Testing and Maintaining
the V24 and V26“in the “V24
and V26 Service Guide”.
Check Calibration of NBP &
Sidestream CO
2
Once a year
See chapter 3 of the CMS
Service Guide “Maintaining
the CMS”.
Check Calibration of all
parameters except NBP &
Sidestream CO
2
Every 2 years
See chapter 3 of the CMS
Service Guide “Maintaining
the CMS”.
Procedure
Frequency
Source of Information
Содержание M1165
Страница 12: ...Responsibility of the Manufacturer xii ...
Страница 62: ...Using an ITE Display 1 40 The CMS and V24 and V26 Patient Monitors ...
Страница 74: ...Attaching the Patient 2 12 Getting Started Getting Started ...
Страница 172: ...Alarm Setup 5 16 Alarm Functions Alarm Functions ...
Страница 228: ...Loading Paper 6 56 Recording Functions ...
Страница 235: ...Admitting a Patient Admit Discharge End Case 7 7 Admit Discharge End Case ICU Mode Adult Pediatric ICU Mode Neonatal ...
Страница 236: ...Admitting a Patient 7 8 Admit Discharge End Case Admit Discharge End Case OR Mode ...
Страница 238: ...Admitting a Patient 7 10 Admit Discharge End Case Admit Discharge End Case endcase tif ...
Страница 274: ...Drug Calculator 8 36 Trends and Calculations Trends and Calculations ...
Страница 299: ...Data Transfer Module Data Transfer 10 3 Data Transfer M1235A CTS DTM CMS CMS V24 V26 ...
Страница 366: ...Accessories and Ordering Information 12 12 Battery Information V24CT and V26CT only Battery Information V24CT and V26CT ...
Страница 388: ...Performance Assurance Checks 13 22 Maintenance Maintenance ...