Troubleshooting
Data Transfer 10-19
D
a
ta
T
ra
n
sf
er
Troubleshooting
Caution
In a centrally networked system, if the previous patient is discharged only
at the monitor and a new patient's data is transferred to the monitor, the
new patient's name and ID will be overwritten by the central station's
broadcast of the previous patient's name. To prevent this situation from
occurring, the previous patient should be discharged from central station
before
the transfer.
The following table outlines the Data Transfer behavior under various
events.
Event
Results
Power Off < 1 min
Transfer will restart automatically
Power Off > 1 min
Transfer is stopped. Collection of new data
must be restarted manually.
Op. Mode Change
Transfer is stopped. Data in module remains
intact.
Config Mode Change
Transfer is stopped. Data in module remains
intact.
Patient Discharge or End
Case
Transfer is stopped. Data in monitor is
deleted. Data in module remains intact.
Содержание M1165
Страница 12: ...Responsibility of the Manufacturer xii ...
Страница 62: ...Using an ITE Display 1 40 The CMS and V24 and V26 Patient Monitors ...
Страница 74: ...Attaching the Patient 2 12 Getting Started Getting Started ...
Страница 172: ...Alarm Setup 5 16 Alarm Functions Alarm Functions ...
Страница 228: ...Loading Paper 6 56 Recording Functions ...
Страница 235: ...Admitting a Patient Admit Discharge End Case 7 7 Admit Discharge End Case ICU Mode Adult Pediatric ICU Mode Neonatal ...
Страница 236: ...Admitting a Patient 7 8 Admit Discharge End Case Admit Discharge End Case OR Mode ...
Страница 238: ...Admitting a Patient 7 10 Admit Discharge End Case Admit Discharge End Case endcase tif ...
Страница 274: ...Drug Calculator 8 36 Trends and Calculations Trends and Calculations ...
Страница 299: ...Data Transfer Module Data Transfer 10 3 Data Transfer M1235A CTS DTM CMS CMS V24 V26 ...
Страница 366: ...Accessories and Ordering Information 12 12 Battery Information V24CT and V26CT only Battery Information V24CT and V26CT ...
Страница 388: ...Performance Assurance Checks 13 22 Maintenance Maintenance ...