Understanding Pulse Oximetry
5-2
Monitoring SpO
2
Understanding Pulse Oximetry
A pulse oximetry sensor sends light through patient tissue to a receiver on the
other side of the sensor. As Figure 5-1 shows, light emitting diodes transmit
red and infrared light through peripheral areas of the body, such as a finger.
Figure 5-1 Pulse Oximetry Sensor
A photodetector positioned opposite the light emitting diodes compares light
absorption before and after pulsation. The amount of light getting through
reflects the blood flow in the arterioles. This measurement of light absorption
during pulsation is translated into an oxygen saturation percentage and an
SpO
2
value is displayed.
For accurate SpO
2
measurements, the following conditions must apply:
z
The patient must have perfusion in that extremity.
z
The light emitter and the photodetector must be directly opposite each
other.
z
All of the light from the emitter must pass through the patient’s tissue.
z
The sensor site should be free of vibration and excessive motion.
z
Power cables should be kept away from the sensor cable and connec-
tor.
LED
Photodetector
Содержание HEARTSTART XLT M3500B
Страница 1: ...H E A R T S TA R T X LT I N S T R U C T I O N S F O R U S E M3500B ...
Страница 2: ......
Страница 3: ...Instructions for Use 0 HDUW6WDUW 7 HILEULOODWRU 0RQLWRU ...
Страница 6: ......
Страница 20: ......
Страница 34: ......
Страница 72: ......
Страница 90: ......
Страница 170: ......
Страница 179: ......
Страница 180: ...M3500 91900 Edition 3 Copyright 2002 Philips Electronics North America Corporation Printed in the U S A October 2002 ...