System Messages
12-6
Troubleshooting
'DWD&DUG)XOO
z
The incident is more than 2 hours in
duration, causing the Data Card to fill.
z
An empty Data Card was not inserted
for the incident, causing the Data Card
to fill sooner.
z
None. A new Data Card can not be
inserted during an incident.
z
Use one Data Card per incident/
patient to decrease the chance of the
card filling.
'DWD&DUG,QWHUUXSWHG
The Data Card is removed during an inci-
dent.
z
None. The Data Card can not be re-
inserted during an incident.
z
Do not remove the Data Card during
an incident.
'DWD&DUG1RW,Q6HUYLFH
The Data Card is inserted while the Heart-
Start XLT is on.
None. A Data Card must be inserted prior
to turning the HeartStart XLT on for the
current patient.
,QFRPSDWLEOH'DWD&DUG
A Data Card other than the M3510A is
inserted.
Use only M3510A Data Cards.
1R'DWD&DUG3UHVHQW
A Data Card is not in the HeartStart XLT.
Turn the HeartStart XLT off and insert a
Data Card prior to the first event for the
patient.
.H\,QDFWLYH
The key pressed is currently inactive (i.e.
and
are inactive in
AED Mode).
Use the appropriate mode for the key.
6WRS3DFHU
is pressed while pacing pulses
are being delivered.
Stop pacing before changing the pacing
mode.
Table 12-2 Momentary Messages
(Continued)
Message
Possible Cause
Corrective Action
Sync
Pacer
Mode
Содержание HEARTSTART XLT M3500B
Страница 1: ...H E A R T S TA R T X LT I N S T R U C T I O N S F O R U S E M3500B ...
Страница 2: ......
Страница 3: ...Instructions for Use 0 HDUW6WDUW 7 HILEULOODWRU 0RQLWRU ...
Страница 6: ......
Страница 20: ......
Страница 34: ......
Страница 72: ......
Страница 90: ......
Страница 170: ......
Страница 179: ......
Страница 180: ...M3500 91900 Edition 3 Copyright 2002 Philips Electronics North America Corporation Printed in the U S A October 2002 ...