Troubleshooting
4-8
Monitoring the ECG
Poor ECG signal
quality
z
The monitoring elec-
trodes are not making
proper contact with the
patient.
z
The monitoring elec-
trodes are outdated or
dried-out.
z
Radio frequency inter-
ference (RFI) is causing
artifact.
z
Check that the monitoring
electrodes are properly
applied. If necessary, pre-
pare the patient’s skin and
apply new electrodes.
z
Check the date code on
the electrodes. Do not
open the electrode pack-
age until immediately
prior to use.
z
Relocate or turn off equip-
ment that may be causing
RFI.
QRS beeper inau-
dible or beeps do
not occur with
each QRS com-
plex.
z
The QRS beeper is con-
figured to Off.
z
The amplitude of the
QRS complex is too
small to detect.
z
Check that the QRS
beeper is configured to
On.
z
Adjust the volume.
z
Adjust the size of the
ECG.
Table 4-5 Troubleshooting Tips
(Continued)
Situation
Cause
Solution
Содержание HEARTSTART XLT M3500B
Страница 1: ...H E A R T S TA R T X LT I N S T R U C T I O N S F O R U S E M3500B ...
Страница 2: ......
Страница 3: ...Instructions for Use 0 HDUW6WDUW 7 HILEULOODWRU 0RQLWRU ...
Страница 6: ......
Страница 20: ......
Страница 34: ......
Страница 72: ......
Страница 90: ......
Страница 170: ......
Страница 179: ......
Страница 180: ...M3500 91900 Edition 3 Copyright 2002 Philips Electronics North America Corporation Printed in the U S A October 2002 ...