Operational Checks
11-2
Maintaining the HeartStart XLT
Operational Checks
The following operational checks are intended to quickly verify the proper
operation of the HeartStart XLT. Perform these checks regularly, at the inter-
vals specified, along with visual inspection of the device and all cables, con-
trols, accessories and supplies. Also regularly check expiration dates of all
supplies, such as multifunction defib electrode pads and monitoring
electrodes.
Before You Begin
Before you run the Shift/System Check, be aware of the following conditions:
z
Do not touch any of the controls on the HeartStart XLT while the
Shift/System Check is running.
z
If a
)DLOXUH
or
6HUYLFH8QLW
message is displayed, or if an unexpected
1RW
7HVWHG
result is displayed, check that the test is set up correctly. Make
sure that:
the paper is in the printer
the test load is attached
a Data Card with enough space is inserted into the HeartStart
XLT
a charged battery is inserted into the HeartStart XLT
Run the Shift/System Check again, ensuring that no one touches any
of the controls on the defibrillator unless prompted to do so.
z
If the Data Card is full, the message
6HUYLFH8QLW
appears at the bottom
of the screen and the message
'DWD&DUG)XOO
appears at the top of the
screen. Replace the Data Card and perform the Shift/System Check
again. If the message
6HUYLFH8QLW
continues to appear, do not use the
device and call for service.
Содержание HEARTSTART XLT M3500B
Страница 1: ...H E A R T S TA R T X LT I N S T R U C T I O N S F O R U S E M3500B ...
Страница 2: ......
Страница 3: ...Instructions for Use 0 HDUW6WDUW 7 HILEULOODWRU 0RQLWRU ...
Страница 6: ......
Страница 20: ......
Страница 34: ......
Страница 72: ......
Страница 90: ......
Страница 170: ......
Страница 179: ......
Страница 180: ...M3500 91900 Edition 3 Copyright 2002 Philips Electronics North America Corporation Printed in the U S A October 2002 ...