6 Patient Alarms and INOPs
80
SpO
₂
NOISY SIGN.
Numeric is replaced by a -?-
INOP tone.
SpO
2
Excessive patient movement or electrical interference is causing
irregular pulse patterns. Try to reduce patient movement or to
relieve the cable strain on the sensor.
SpO
₂
NON-PULSAT.
Numeric is replaced by a -?-
INOP tone.
SpO
2
Check the perfusion at measurement site. If necessary, stimulate
circulation or change measurement site. If the INOP is due to
noninvasive blood pressure measurement on the same limb, wait
until the measurement is finished.
SpO
₂
NO SENSOR
Numeric is replaced by a -?-
INOP tone.
SpO
2
Make sure the SpO
2
sensor is connected. If the INOP persists,
try another adapter cable and sensor. If you silence this INOP,
the measurement will be switched off.
SpO
₂
POOR SIGNAL
Label is preceded by a
? (questionable numeric)
SpO
2
The signal condition of the SpO
2
measurement is poor and
measurement accuracy may be compromised.
SpO
₂
PULSE?
Numeric is replaced by a -?-
INOP tone.
SpO
2
The detectable pulsations of the SpO
2
signal are outside the
specified pulse rate range.
SpO
₂
SEARCHING
Numeric unavailable
SpO
2
SpO
2
is analyzing the patient signal to derive Pulse, and SpO
2
values. Please wait until the search analysis is complete.
SpO
₂
SENSOR MALF
Numeric is replaced by a -?-
INOP tone.
SpO
2
The SpO
2
sensor or adapter cable is faulty. Try another adapter
cable and sensor. If the INOP persists, contact your service
personnel.
SpO
₂
SENSOR OFF
Numeric is replaced by -?-
INOP tone.
SpO
2
The SpO
2
sensor is not properly applied to the patient. Apply
the sensor following the instructions supplied by the
manufacturer.
SpO
₂
UNKN.SENSOR
Numeric is replaced by a -?-
SpO
2
The connected sensor or adapter cable is not supported by the
SpO
2
measurement. Use only specified sensors and cables.
SpO
₂
UPGRADE
Label is replaced by a -?-, or
numeric is unavailable.
SpO
2
The SpO
2
measurement is currently in UPGRADE mode.
Monitoring is not possible in this mode.
TimeExpired:NST
Monitor
The time has expired for the NST timer. Clearing the timer
clears the INOP.
Toco EQUIP MALF
INOP tone
Toco
There is a problem with the Toco hardware. Contact your
service personnel.
INOP Message, Indication
Source
What to do
Содержание Avalon FM20
Страница 2: ......
Страница 8: ...8 ...
Страница 56: ...3 Basic Operation 56 ...
Страница 72: ...5 Alarms 72 ...
Страница 82: ...6 Patient Alarms and INOPs 82 ...
Страница 86: ...7 Admitting and Discharging 86 ...
Страница 90: ...8 Non Stress Test Timer 90 ...
Страница 120: ...11 Monitoring Twin FHRs 120 ...
Страница 126: ...12 Monitoring Triple FHRs 126 ...
Страница 136: ...15 Monitoring Uterine Activity Internally 136 ...
Страница 142: ...16 Monitoring FHR Using DECG 142 Testing DECG Mode Refer to the monitor s Service Guide ...
Страница 154: ...18 Monitoring SpO2 154 ...
Страница 164: ...20 Printing the ECG Waveform 164 ...
Страница 166: ...21 Paper Save Mode for Maternal Measurements 166 ...
Страница 170: ...22 Recovering Data 170 ...
Страница 174: ...23 Care and Cleaning 174 ...
Страница 188: ...25 Accessories and Supplies 188 ...
Страница 216: ...26 Specifications and Standards Compliance 216 ...
Страница 221: ...Part Number 453564290161 Printed in Germany 08 11 453564290161 ...