18 Monitoring SpO2
153
Alarm Delays
There is a delay between a physiological event at the measurement site and the corresponding alarm at
the monitor. This delay has two components:
•
The general system delay time is the time between the occurrence of the physiological event and
when this event is represented by the displayed numerical values. This delay depends on the
algorithmic processing and the configured averaging time. The longer the averaging time
configured, the longer the time needed until the numerical values reflect the physiological event.
•
The time between the displayed numerical values crossing an alarm limit and the alarm indication
on the monitor. This delay is the combination of the configured alarm delay time plus the general
system alarm signal delay time.
Adjusting the SpO2 Alarm Limits
In the
Setup SpO
₂
menu:
•
Select
High Limit
then choose the upper alarm limit.
•
Select
Low Limit
then choose the lower alarm limit.
Adjusting the Desat Limit Alarm
The Desat alarm is a high priority (red) alarm notifying you of potentially life threatening drops in
oxygen saturation.
1
In the
Setup SpO
₂
menu, select
Desat Limit
.
2
Adjust the limit.
Adjusting the Pulse Alarm Limits
See “Adjusting the Heart Rate / Pulse Alarm Limits” on page 160.
Setting Up Tone Modulation
If tone modulation is on, the QRS tone pitch lowers when the SpO
2
level drops. Remember, the QRS
tone is derived from either heart rate (from MECG) or pulse (from SpO
2
) depending on which is
currently displayed (see “Priority for Maternal Heart / Pulse Rate” on page 155).
NOTE
Pulse from Toco MP does not provide a QRS tone.
Setting the QRS Volume
In the
Setup SpO
₂
menu, select
QRS Volume
and set the appropriate QRS tone volume.
Содержание Avalon FM20
Страница 2: ......
Страница 8: ...8 ...
Страница 56: ...3 Basic Operation 56 ...
Страница 72: ...5 Alarms 72 ...
Страница 82: ...6 Patient Alarms and INOPs 82 ...
Страница 86: ...7 Admitting and Discharging 86 ...
Страница 90: ...8 Non Stress Test Timer 90 ...
Страница 120: ...11 Monitoring Twin FHRs 120 ...
Страница 126: ...12 Monitoring Triple FHRs 126 ...
Страница 136: ...15 Monitoring Uterine Activity Internally 136 ...
Страница 142: ...16 Monitoring FHR Using DECG 142 Testing DECG Mode Refer to the monitor s Service Guide ...
Страница 154: ...18 Monitoring SpO2 154 ...
Страница 164: ...20 Printing the ECG Waveform 164 ...
Страница 166: ...21 Paper Save Mode for Maternal Measurements 166 ...
Страница 170: ...22 Recovering Data 170 ...
Страница 174: ...23 Care and Cleaning 174 ...
Страница 188: ...25 Accessories and Supplies 188 ...
Страница 216: ...26 Specifications and Standards Compliance 216 ...
Страница 221: ...Part Number 453564290161 Printed in Germany 08 11 453564290161 ...