3 Basic Operation
34
CAUTION
This resets all settings to factory defined values, but be aware that some values will differ from those
with which the monitor was originally shipped from the factory (recorder speed and paper scale type
will need to be corrected, for instance). After loading the Factory Default, please check the settings
and, if necessary, change them to the settings you normally use.
You can use the Factory Default as the basis for producing your User Default. See the
Service Guide
for
details.
Global Settings
General monitor configuration settings are stored in the Global Settings. These include settings for line
frequency, QRS type and whether the monitor is automatically reset to the User Default after a power
interruption of more than one minute. You can change the Global Settings in Configuration Mode.
Changing Measurement Settings
Each measurement has a setup menu in which you can adjust all of its settings. You can enter a setup
menu:
•
via the measurement numeric - select the measurement numeric on the screen to enter its setup
menu. For example, to enter the
Setup FHR1
menu, select the
FHR1
(fetal heart rate 1) numeric.
•
via the
Main Setup
SmartKey - if you want to setup a measurement when the measurement is
switched off, use the
Main Setup
SmartKey and select
Measurements
. Then select the
measurement name from the pop-up list. With this SmartKey you can access any setup menu in
the monitor.
This guide always describes the entry method using the measurement's setup menu. You can use the
method you prefer.
Switching the Noninvasive Blood Pressure Measurement On and
Off
The noninvasive blood pressure measurement is the only measurement for which you can manually
switch on and off. To do this:
1
Enter the noninvasive blood pressure measurement's setup menu.
2
Select
NBP
to toggle between on and off. The screen display indicates the active setting.
Changing Monitor Settings
To change monitor settings such as brightness, or touch tone volume:
1
Enter the
Main Setup
menu by selecting the SmartKey
.
2
Select the setting you want to change, or select
User Interface
to enter a submenu where you can
change user interface settings.
Содержание Avalon FM20
Страница 2: ......
Страница 8: ...8 ...
Страница 56: ...3 Basic Operation 56 ...
Страница 72: ...5 Alarms 72 ...
Страница 82: ...6 Patient Alarms and INOPs 82 ...
Страница 86: ...7 Admitting and Discharging 86 ...
Страница 90: ...8 Non Stress Test Timer 90 ...
Страница 120: ...11 Monitoring Twin FHRs 120 ...
Страница 126: ...12 Monitoring Triple FHRs 126 ...
Страница 136: ...15 Monitoring Uterine Activity Internally 136 ...
Страница 142: ...16 Monitoring FHR Using DECG 142 Testing DECG Mode Refer to the monitor s Service Guide ...
Страница 154: ...18 Monitoring SpO2 154 ...
Страница 164: ...20 Printing the ECG Waveform 164 ...
Страница 166: ...21 Paper Save Mode for Maternal Measurements 166 ...
Страница 170: ...22 Recovering Data 170 ...
Страница 174: ...23 Care and Cleaning 174 ...
Страница 188: ...25 Accessories and Supplies 188 ...
Страница 216: ...26 Specifications and Standards Compliance 216 ...
Страница 221: ...Part Number 453564290161 Printed in Germany 08 11 453564290161 ...