IEC 60601-1 classification
Type of Protection
Against Electric Shock
Class II Equipment
Degree of Protection
Against Electric Shock
Type BF Applied Part
Degree of Protection
Against Ingress of Water
Device: Drip Proof, IP22
First characteristic numeral - 2 - Protection against ingress of solid
foreign objects
≥
12.5 mm diameter.
Explanation: Protected against access to hazardous parts with a finger
and protected against solid foreign objects of 12.5 mm diameter and
greater.
Second characteristic numeral - 2 - Protection against ingress of water
with harmful effects dripping (15° tilted).
Explanation: Protected against vertically falling water drops when
enclosure tilted up to 15°.
Mode of Operation
Continuous
Electrical
AC Power Consumption: 100–240
VAC, 50/60 Hz, 2.0-1.0 A
Water capacity
325 ml (11 oz.) at recommended
water level
Heater plate
Max Temperature: 68°C (154°F)
Humidity
Humidity
min
Output:
≥
12 mg H
2
O/L
Measured at expected leak across
therapy pressures , 17 °C– 35°C,
15% RH for all compatible tubes.
Maximum delivered gas temperature
<43°C (109.4°F)