background image

13 

 

PM5A2 
 

6.2.3  TIMER Mode 

 

The use of 

 key allows the 

PM5A2

 setting for the performance of a pre-set - duration test. When this mode is 

selected, the display shows the programmed time. Use 

 and 

  keys to define the duration of the tests in 1 

minute, 3 minutes, 10 minutes, 30 minutes or 90 minutes. 

 

 

 
 

6.2.4  Pass / Fail Test mode 

Press 

 

key in order to determine the lower insulation limit for type “Pass / Fail” test. Select this value using 

 

and 

 keys. Possible values are 10 

MΩ, 100 MΩ, 1 GΩ

or 10 

GΩ. 

 

 

 
 

During a Pass/ Fail test, the 

PM5A2

 will indicate when the insulation resistance is lower than the programmed limit, 

both with an intermittent beep and the 

 key led flashing.  The 

 key LED will remain flashing until the end of 

tests, or until the measurement of the resistance value is greater than the programmed limit. 
 

 

 

Содержание PM5A2

Страница 1: ...ion Tester User s Manual Phenix Technologies Inc 75 Speicher Drive Accident Maryland 21520 Tel 1 301 746 8118 Fax 1 301 895 5570 E mail info phenixtech com Rev 3 2017 GU 1547RB 2016 Phenix Technologie...

Страница 2: ...sure that there are not any voltage difference between the points to which the equipment will be connected to neither between them and ground The panel terminals and connectors of the equipment must s...

Страница 3: ...Fail Test mode 13 7 How to perform tests 14 7 1 Measurement of the Dielectric Absorption Index DAI 15 7 2 Measurement of the Polarization Index PI 15 8 Other functions 16 8 1 Backlight 16 8 2 Filter...

Страница 4: ...olarization and Dielectric absorption index automatic measurement Measurement of Leakage Current and Capacitance TIMER to program resistance test time Limit to allow making test type Pass Fail with pr...

Страница 5: ...Functions 1 Voltage Output Terminal V 2 Zero Reference Terminal R 3 Guard G Terminal 4 Display Screen 5 High Voltage LED 6 Keyboard 7 On Off Key 8 USB Communication Port 9 Power Supply Input 10 Print...

Страница 6: ...t SVT Polarization Index PI and Dielectric Absorption Index DAI Backlight activates the display light Capacitance displays the capacitance value Activated enables the programming of the Operation Mode...

Страница 7: ...tes enables programming of the limit for the Pass Fail test Indicates when the measured resistance is lower than programmed limit Start Start test Indicates that the test is being executed Stop End of...

Страница 8: ...s connected during one second Steady red light Battery under charge Flashing red light The battery was not successfully charged It denotes some trouble in the battery charging process Steady green lig...

Страница 9: ...omatically enters in the voltmeter mode and displays the circuit voltage The circuit to be tested must be de energized to avoid interferences in the measurement The equipment will block the start of m...

Страница 10: ...ment potential setting is fulfilled If the measurement is performed between two parts which are not grounded for example between two phase conductors in a three phase cable the equipment GUARD termina...

Страница 11: ...both the pre programmed voltage selection and the and keys that increase or decrease the value of the step voltage test for 25 V 100 V or 500 V depending on the selected voltage adjustment key As lon...

Страница 12: ...s each minute until reaching the programmed voltage At each stage the PM5A2 measures the resistance before advancing towards the following step The use of voltage adjusting keys determines the value o...

Страница 13: ...Pass Fail Test mode Press key in order to determine the lower insulation limit for type Pass Fail test Select this value using and keys Possible values are 10 M 100 M 1 G or 10 G During a Pass Fail te...

Страница 14: ...played with its corresponding unit and the analog indication will start by bargraph In order to end the test press the key At that moment the last measured values will remain frozen in the PM5A2 displ...

Страница 15: ...ent in transformers motors generators etc D A I R R 6 0 s e c o n d s 3 0 s e c o n d s 7 2 Measurement of the Polarization Index PI When pressing the key during the test the Polarization Index PI val...

Страница 16: ...this function connect the test points and turn on the unit The measured value will be exhibited automatically in the display AC DC 10V up to 1000Vr m s 10V up to 1000V Precision 5 of the reading 3 dig...

Страница 17: ...be replaced by the newest ones To avoid loss of data always download the internal memory after finishing the measurements 8 7 Battery status check Press and hold the key in order to check the battery...

Страница 18: ...e device If no suitable driver is automatically found then insert the CD ROM supplied with the equipment in the PC run the executable usb install exe and click Install 9 1 MegaLogg2 software This soft...

Страница 19: ...order to print it completely including the heading CAUTION Don t pull the paper The printer can be easily damaged This printer uses 56mm wide thermal paper which comes in a 30mm diameter reel Pull the...

Страница 20: ...t 1nA up to 3mA 10 of the reading 3 digits Capacitance measurement 50nF up to 10 F 500V 50nF up to 5 F 1 000V 30nF up to 2 F 2 500V 30nF up to 1 F 5 000V Precision 10 of the reading 3 digits Short cir...

Страница 21: ...x IP54 with closed lid Safety In accordance with IEC 61010 1 Electromagnetic compatibility E M C In accordance with IEC 61326 1 Electromagnetic irradiation immunity In accordance with IEC 61000 4 3 El...

Страница 22: ...er micro ammeter The unknown resistance Rx is connected between V and R terminals Its value determines the current passing through the circuit which in turn is indicated by the micro ammeter The value...

Страница 23: ...its influence is reduced during reading Through resistance R2 circulates a current which is not passing through the meter and consequently does not affect the reading In fact current through R2 origin...

Страница 24: ...s Manual are correct If there are concerns or comments as you have used this information Phenix Technologies appreciates any feedback Unit Serial Number Sect Page s Comment Please return to Phenix Te...