TROUBLESHOOTING
If the unit malfunctions, check out the following before calling for service. In the case of no refrigeration
or poor refrigeration, transfer the stored items to another refrigerator or freezer before checking out.
Malfunction Check/Remedy
If nothing operates even
when switched on
□
The unit is not connected to the power supply or capacity of power
source is not enough.
□
There is a power failure, the fuse is blown, or the circuit breaker is
activated.
When the unit does not
accept change of setting
□
The key lock function is set in ON mode.
When alarm is activated
□
On start-up
The temperature in the unit does not match set value.
□
In use
The door was kept opened for long time.
The set value was changed.
The containers of high temperature (load) were put in the unit. In
these cases, alarm is removed automatically by running the unit for
several hours.
When unit does not get
cold enough
□
The air exhaust vent is blocked by refrigerator contents.
□
A large amount of warm product was put in the unit.
□
There is any heat sources in the unit.
□
The door is opened frequently.
□
The ambient temperature is too high.
□
The unit is in direct sunlight.
□
The door is not securely closed.
□
The door seal is damaged or foreign substance inserted between
the door gaskets.
Note:
If the malfunction is not eliminated after checking the above items, or the malfunction is not shown in the
above table, contact our sales representative or agent.
DISPOSAL OF UNIT
WARNING
If the unit is to be stored unused in an unsupervised area for an extended period
ensure that children
do not have access and doors cannot be closed completely.
The disposal of the unit should be accomplished by appropriate personnel.
Always remove
doors
to prevent accidents such as suffocation.
Decontamination of unit
Before disposing the unit with biohazardous danger, decontaminate the unit to the extent possible by the
user.