12
INSTALLATION
1. Remove the packaging materials and tapes
Remove all transportation packaging materials and
tapes. Open the doors and ventilate the unit. If the
outside panels are dirty, clean them with a diluted
neutral dishwashing detergent. (Undiluted detergent
can damage the plastic components. For the dilution,
refer to the instruction of the detergent.) After the
cleaning with the diluted detergent, always wipe it off
with a wet cloth. Then wipe off the panels with a dry
cloth.
Note:
Remove the cable tie banding the power supply cord.
Prolonged banding may cause the corrosion of the cord
coating.
2. Adjusting the leveling foot
Extend the leveling feet by rotating them
counterclockwise to contact them to the floor.
Ensure the unit is level. (Fig.1)
3. Fixing the unit
Two fixtures are attached to the rear of the frame. Fix
the frame to the wall with these fixtures and rope or
chain. (Fig.2)
4. Ground (earth)
WARNING
Use a power supply outlet with ground (earth)
to prevent electric shock. If the power supply outlet is
not grounded, it is necessary to install a ground by qualified engineers.
Never ground the unit through a gas pipe, water main, telephone line or lightning rod.
Such
grounding may cause electric shock in the case of an incomplete circuit.
Note:
If an instrument requiring a power source is to be placed inside the cabinet, the cable can be lead through
the access port on the back side of the cabinet. After using of the port, a rubber cap and insulation
should be replaced to seal the access port. Failure to do this can affect the temperature uniformity
inside the cabinet and lead to condensation on the outside of the access port.
Leveling foot
Fixture
Fig.1
Fig.2