ALARM TEMPERATURE SETTING
Table 5
Operation
Key operated
Indication after operation
1
-----
The current refrigerator temperature
is displayed.
2
Press and hold the numerical value
shift key for about 5 seconds.
The first digit of the temperature
display blinks.
3
Set the first digit to 2 with the
numerical value shift key. (Note 2)
Pressing the key shifts up the figure
of the current digit.
4 Press the set key.
SET
The current set value is displayed
and the first digit of the temperature
display flashes.
5
Set to -03 by using digit shift key
and numerical value shift key.
Pressing the key leads the change of flashed
digit.
Pressing the key shifts up the figure
of the current digit.
6 Press the set key.
SET
The value is stored in memory and
the current refrigerator temperature
is displayed.
Note 2:
For the freezer, set the first digit to 4 (F04).
Note:
If the ambient temperature is fairly high, the alarm lamp is flashed, temperature display is flashed, and
the alarm buzzer sounds at the time of initial start up. The alarm is canceled automatically when the
chamber temperature is decreased.
Always close the door firmly. The door check lamp is lit when the door is open. The alarm buzzer
sounds two minutes after door opening. The alarm buzzer can be canceled automatically when the door
is closed.