   START-UP OF UNIT

 

 
Use the following procedure to start trial operation or actual operation of the unit. 

 
1.

 Check the power switch is off, battery switch is off and sub-heater switch is off. 

 
2.

 Connect the power supply cord to the dedicated outlet with appropriate rating.  Then turn on the 

power switch. 

 
3.

 Turn on the battery switch. 

Note: 

The battery needs to be replaced about every 3 years.    Contact Our sales representative or agent for 
battery replacement. 
 

4.

 The alarm buzzer sometimes operates.    In this case, stop the buzzer by pressing the alarm buzzer 

stop key (BUZZER).

 

 
5.

 Set the chamber temperature to the desired temperature. 

 
6.

 Allow the chamber temperature to fall to the desired temperature.    Check the chamber temperature 

on the digital temperature indicator.

 

 
7.

 Press the alarm test key (ALARM TEST) and check that the alarm lamp blinks and alarm buzzer 

activates. 

 
8.

 Begin slowly placing items into the chamber to minimize the temperature rise. 

 

CAUTION

 

Do not put too many warm articles into a freezer compartment before enough operating.    Put items in a 
few at a time after the freezer compartment temperature has cooled to at least -20

o

C.

 

 
Start-up of unit: 

The unit is programmed to repeatedly turn the compressor on and off in order to make its startability 
better when you start the unit at low ambient temperature (about 25

o

C or lower).  However, the unit 

would start the operation without turning the compressor on and off repeatedly when you install the 
cooled unit at high ambient temperature and it causes the unit would have the starting current running 
with longer time. 
The unit should be exposed to ambient temperature sufficiently, 4 hours at least to avoid the breaker from 
trapping. 
 

    Operation after power failure

 

 
The memory (chamber temperature setting and alarm temperature setting) is backed-up by nonvolatile 
memory.    Accordingly, the freezer resumes the operation with setting before power failure. 

 

CAUTION

 

The start-up after power failure is subject to adverse affection such as voltage drop because all electrical 
appliance start to operate simultaneously.    Always check the running status.

 

 

Содержание MDF-U443

Страница 1: ...U443 Please read the operating instructions carefully before using this product and keep the operating instructions for future use See page 35 for model number Operating Instructions Biomedical Freez...

Страница 2: ......

Страница 3: ...RM P 17 SETTING OF ALARM RESUME TIME P 18 SETTING OF DOOR ALARM DELAY TIME P 19 SETTING OF COMPRESSOR DELAY TIME P 20 DEFROSTING P 21 Automatic defrosting P 21 Manual defrosting P 21 Sub heater switch...

Страница 4: ...ating instructions are subject to change without notice for improvement of performance or function Contact our sales representative or agent if any page of the operating instructions is lost or the pa...

Страница 5: ...her person Precautions are illustrated in the following way WARNING Failure to observe WARNING signs could result in a hazard to personnel possibly resulting in serious injury or death CAUTION Failure...

Страница 6: ...wer supply outlet is not grounded it will be necessary to install a ground by qualified engineers Never ground the unit through a gas pipe water main telephone line or lightning rod Such grounding may...

Страница 7: ...e electric shock or fire When removing the plug from the power supply outlet grip the power supply plug not the cord Pulling the cord may result in electric shock or fire by short circuit Disconnect t...

Страница 8: ...r improper insertion may cause a heat or ignition Never store corrosive substances such as acid or alkali in this unit if the container cannot be sealed These may cause corrosion of inner components o...

Страница 9: ...rs on the chamber wall side It cannot be cooled down when there is much amount of frost Personal injury Sample damage Injury or damage Human inserted things Failure of equipment Interior When insertin...

Страница 10: ...p to 2000 m Ambient temperature 5 o C to 40 o C Maximum relative humidity 80 for temperature up to 31 o C decreasing linearly to 50 relative humidity at 40 o C Mains supply voltage fluctuations up to...

Страница 11: ...9 FREEZER COMPONENTS 12 1 2 4 3 10 7 5 8 9 13 13 11 15 16 17 21 20 22 18 19 6 14 13 Switch cover...

Страница 12: ...2 Shelf Shelf location can be freely 13 Access port left side rear side This is used for leading the measuring cable from the chamber to the outside 14 Fixture rear side 15 Lamp Lights up when door is...

Страница 13: ...in the setting mode causes the changeable digit to shift Key lock setting mode is led by pressing this key for more than 5 seconds in the temperature display mode Refer to page 15 for the key lock fu...

Страница 14: ...n or noise WARNING Install the unit on a sturdy floor If the floor is not strong enough or the installation site is not adequate this may result in injury from the unit falling or tipping over Select...

Страница 15: ...ot Stretch the leveling feet by rotating them to contact them to the floor Ensure the unit is installed horizontally Fig 1 3 Fixing the unit Two fixtures are attached to the rear of the frame Fig 2 Fi...

Страница 16: ...N Do not put too many warm articles into a freezer compartment before enough operating Put items in a few at a time after the freezer compartment temperature has cooled to at least 20oC Start up of un...

Страница 17: ...bient temperature at 35 o C The temperature set mode returns to the temperature display mode automatically when 90 seconds has passed without any key operation KEY LOCK FUNCTION This unit is provided...

Страница 18: ...inutes The procedure in table 3 shows the sequence to set the delay time to 10 minutes The delay time is set to 15 minutes at the factory Table 3 Setting procedure for alarm delay time change from 15...

Страница 19: ...to 1 When pressed the figure of settable digit change 4 Press set key SET The current setting is displayed and the first digit is flashed 5 Scroll the figure to 010 by using the numerical value shift...

Страница 20: ...5 with the numerical value shift key and the digit shift key When pressed the figure of settable digit changes The settable digit is shifted 4 Press set key SET The current setting is displayed The se...

Страница 21: ...mber temperature is displayed 2 Press numerical value shift key for 5 seconds The first digit is flashed 3 Set the figure to F04 with the numerical value shift key When pressed the figure of settable...

Страница 22: ...ation 1 The current chamber temperature is displayed 2 Press numerical value shift key for 5 seconds The first digit is flashed 3 Set the figure to F05 with the numerical value shift key When pressed...

Страница 23: ...to start the defrosting process the automatic defrosting of 12 hour cycle is started at this point Manual defrosting In case of manual defrosting perform key operations as follows 1 Press the defrost...

Страница 24: ...ture is displayed flashed Intermittent tone with 15 minutes delay Remote alarm with 15 minutes delay Low temperature alarm If the chamber temperature is lower than the temperature at which the low tem...

Страница 25: ...Alarm lamp is flashed E08 and chamber temp are displayed alternately Intermittent tone Remote alarm Battery switch check When the battery switch is OFF during alarm test Alarm lamp is flashed E09 is f...

Страница 26: ...of the detergent After the cleaning with the diluted detergent always wipe it off with a wet cloth Then wipe off the cabinet or accessories with a dry cloth Never pour water onto or into the unit Doi...

Страница 27: ...ng dust and other foreign substances accumulated on the condenser filter surface wash it with clean water and dry it naturally 5 After cleaning replace the condenser filter 6 Replace and fix the left...

Страница 28: ...tch battery switch sub heater switch and disconnect the power supply plug 2 As shown in the Fig 1 remove the unit cover by unscrewing 4 fixing screws with a screw driver 3 Pull out the unit cover and...

Страница 29: ...hours The cooling is poor The environmental temperature is too high The door is not shut tightly The filter is clogged The freezer is in the direct sunlight There is any heating source near the freeze...

Страница 30: ...wers the product you have purchased Please call 1 800 8 BATTERY for information on how to recycle this battery L appareil que vous vous tes procur est aliment par une pile au nickel m tal hydrure NiMH...

Страница 31: ...the chemical involved German Entsorgung von Altger ten und Batterien Nur f r die Europ ische Union und L nder mit Recyclingsystemen Dieses Symbol auf den Produkten der Verpackung und oder den Begleitd...

Страница 32: ...r sur les piles pictogramme du bas Si ce pictogramme est combin avec un symbole chimique il r pond galement aux exigences pos es par la Directive relative au produit chimique concern Spanish Eliminaci...

Страница 33: ...Neste caso estar em conformidade com o estabelecido na Diretiva referente aos produtos qu micos em causa Italian Smaltimento di vecchie apparecchiature e batterie usate Solo per Unione Europea e Nazi...

Страница 34: ...l Onderstaand symbool Dit symbool kan in combinatie met een chemisch symbool gebruikt worden In dit geval volstaan de eisen die zijn vastgesteld in de richtlijnen van de desbetreffende chemische stof...

Страница 35: ...der contact our sales representative or agent For the usage of the temperature recorder also refer to an installation sheet enclosed with the temperature recorder Main specifications of temperature re...

Страница 36: ...R 134a HFC 134a high stage side R 404A low stage side Temperature control Microprocessor control system Temperature display Digital display Temperature sensor Thermister sensor Alarm High temperature...

Страница 37: ...center of chamber ambient temperature 35o C no load Temperature control range 15o C to 40o C Rated voltage AC 230 V 240 V Rated frequency 50 Hz Amperage 4 1 A 4 5 A Rated power consumption 650 W 680 W...

Страница 38: ...the unit Unit interior No contamination Decontaminated Contaminated Others 3 Instructions for safe repair maintenance disposal of the unit a The unit is safe to work on b There is some danger see belo...

Страница 39: ......

Страница 40: ...rating Instructions Printed in Japan LDCL034200 0 S0418 0 2018 04 01 1 1 1 Sakada Oizumi machi Ora gun Gunma 370 0596 Japan EU countries only PHC Corporation 2018 Nijverheidsweg 120 4879 AZ Etten Leur...

Отзывы: