19
ALARM TEMPERATURE SETTING
Table 3 Procedure for setting high temperature alarm (Top compartment)
Description of operation
Key operated
Indication after operation
1
―
―
The current chamber temperature is
displayed.
2
Press numerical value shift key for
5 seconds.
The display changes to F00 and
the first digit blinks.
3
Set the figure to F01 by using
digit shift key and numerical value
shift key
When pressed, the figure of
settable digit changes.
4 Press set key.
SET
The current setting is displayed and
the first digit is flashed.
5
Set the figure to 005 by using
digit shift key and numerical value
shift key
When pressed, the settable digit is
shifted.
When pressed, the figure of
settable digit changes.
6 Press set key.
SET
Alarm temperature is memorized and
the current chamber temperature is
displayed.
• The set mode returns to the temperature display mode automatically when 90 seconds has passed
without any key operation. In this case, any setting before pressing set key (SET) is not memorized.
Table 4 Procedure for setting low temperature alarm (Top compartment)
Description of operation
Key operated
Indication after operation
1
―
―
The current chamber temperature is
displayed.
2
Press numerical value shift key for
5 seconds.
The display changes to F00 and
the first digit blinks.
3
Set the figure to F02 by using
digit shift key and numerical value
shift key
When pressed, the figure of
settable digit changes.
4 Press set key.
SET
The current setting is displayed and
the first digit is flashed.
5
Set the figure to -05 by using
digit shift key and numerical value
shift key
When pressed, the settable digit is
shifted.
When pressed, the figure of
settable digit changes.
6 Press set key.
SET
Alarm temperature is memorized and
the current chamber temperature is
displayed.
• The set mode returns to the temperature display mode automatically when 90 seconds has passed
without any key operation. In this case, any setting before pressing set key (SET) is not memorized.
Содержание MDF-MU539D
Страница 39: ......