16
START-UP OF UNIT
Use the following procedure to start trial operation or actual operation of the unit.
1.
In the empty state, power supply plug the plug into a power supply outlet. Confirm that the temperature
display on the control panel is blinking.
2.
Turn on the battery switch.
<Important>
Always turn the battery switch on when the unit is operating to ensure that the power-failure alarm is
working.
3.
Set the desired chamber temperature. See page 17 for the temperature setting.
4.
Check that the chamber temperature reaches the desired temperature.
5.
Make sure that the alarm lamp flashes and the buzzer sounds by pressing the alarm test key (ALARM
TEST). The remote alarm is also operated. E09 is displayed on the digital temperature indicator and
buzzer sounds if the battery switch is off. Make sure to turn on the battery switch.
6.
After confirming the above, you can put articles into the biomedical freezer chamber in a small batch to
prevent the temperature rise.
Note:
■ When starting the operation of the biomedical freezer for the first time, the alarm lamp (ALARM) lights.
When the chamber temperature reaches around the set temperature, then the alarm lamp goes out (The
remote alarm is activated).
■ If the battery switch is turned on before turning on the power of the biomedical freezer, the temperature
alarm is activated and the buzzer sounds and the remote alarm is also activated. Check that the battery
switch is off before turning on the biomedical freezer.