18
CAUTIONS FOR USING EQUIPMENT IN CHAMBER
●
Use the inner outlet for an equipment used in the chamber
For power source of a heat emitting equipment used in the chamber such as a shaker or a bottle roller,
use the electrical outlet equipped in the product. In case of that the chamber temperature rises too much
due to this heat emission, when it reaches the preset temperature of over heating protection, electricity
through the outlet is cut automatically to avoid over heating in the chamber. For this reason, avoid using
equipment in the chamber that take power source from any other than the outlet equipped in the product,
since it may be dangerous.
●
If there is too much heat emission from the equipment in the chamber
The chamber temperature may deviate very much from the setting temperature when a heat generator
(heater, motor, etc.) of the equipment is energized in the chamber.
●
When operating an equipment in the chamber, use at "No humidifying mode" (refer to page 23).
High humidity may cause a failure, a current leakage or a fire to an equipment in the chamber.
●
When operating an equipment in the chamber, make sure there is no water in the water
reservoir.