10
INCUBATOR COMPONENTS
1.
Outer door:
Sticks to frame with magnetic seal. Door heater is installed in the door panel. The door
window is double glass.
2. Leveling foot:
Screw type for adjusting the height. Adjust the foot so that the unit can be level.
3. Tray:
Can be pulled toward you. Make sure to set the bottom tray to the bottom position of side duct
with tray support (This may cause condensation on bottom cover by varying air flow inside chamber).
4. Tray support:
Can be removed by lifting the front side and pulling toward you.
5. Side duct:
Flow path for circulating air. Removable.
6. Fan (behind of ceiling)
7. Sample air outlet:
This also functions as an internal gas outlet. Normally, cover this outlet with the
sample air outlet cap.
8. Sample air outlet cap:
Always attach this cap except at the time of using of sample air outlet.
9. Door switch:
Detects the door opening/closing and stops the circulating fan and electromagnetic
valve for CO
2
when door is open. UV lamp is also deactivated by door opening (when an optional UV
system set MCO-80UVS is installed).
10. Inside power outlet:
AC 115 V and up to 3 A in total. It is possible to use on the condition that inner
temperature below 40
o
C and humidity below 50%R.H. When humidifying chamber, bottom cover must be
placed properly. At the first time to operate, press the reset button between two inside power outlets.
11. Access port:
When not in use, cap them with two attached silicon caps on outside.
12. Connecting port for CO
2
gas pipe (rear side):
When an optional component MCO-80GC (gas
auto changer) is installed, both A and B are available. If MCO-80GC is not used, only A is available.
Refer to page 17 for gas cylinder connection. Ensure that the gas pressure is set at 0.1 MPa(G) (1.0
kgf/cm
2
(G), 14.5 psi(G)). Refer to page 36 for gas auto changer.
13. Power switch:
Main switch of the unit (ON-“I”, OFF-“O”). Also functions as an over-current
breaker.
14. Power switch cover:
Power switch is covered by a power switch cover to prevent the accidental
push. To turn on or off the switch, remove a power switch cover by loosening the screw.
15. Remote alarm terminal:
Refer to page 14.
16. Outer power outlet:
AC 115 V, and up to 3 A in total.
17. Bottom cover:
Prevents UV light being exposed to the chamber. Lift to remove for cleaning. See
page 29 for details.
18. Water supply inlet cover:
When filling the water reservoir, remove a screw and rotate the cover.
19. Water supply inlet:
Usually closed with water supply inlet cover.
20. Drain cock:
Open the drain cock to drain water from the water reservoir
.
21. Drain outlet:
Insert the drain outlet into the attached drain tank installed under the unit.
22. Drain tank
23. Front panel lower cover:
Detach it when draining water from the water reservoir. See to page 29
for details.