16
INSTALLATION SITE
■
Altitude up to 1000 m
It is necessary to replace the outer glass door with specific one when the unit is used at an altitude of
1000 m or higher. If the unit needs to be airfreighted, consult with our sales representative or agent.
INSTALLATION
1. Remove the packaging materials and tapes
Remove all transportation packaging materials and tapes. Open the doors and ventilate the unit. If the
outside panels are dirty, clean them with a thinned neutral detergent and wipe it up with a wet cloth.
Note:
Remove the cable tie banding the power supply cord. Prolonged banding may cause the corrosion of
the cord coating.
2. Adjust the leveling feet
Extend the leveling feet by rotating them counterclockwise to contact them to the floor. Ensure the unit
is level.
3. Fix the unit
Two fixtures are attached to the rear of the frame. Fix the frame to the wall with these hooks and rope or
chain.
4. Ground (earth)
WARNING
Use a power supply outlet with ground (earth)
to prevent electric shock. If the power supply outlet is
not grounded, it is necessary to install a ground by qualified engineers.
Never ground the unit through a gas pipe, water main, telephone line or lightning rod.
Such
grounding may cause electric shock in the case of an incomplete circuit.
●
If not used
When the unit is not used, drain the water in the water reservoir and remove any moisture in the chamber
completely. Check that the chamber is completely dry before closing the door.
●
Before moving the unit
Empty the water reservoir completely before moving the unit. Spilled or splashed water may cause
current leakage or electric shock.