PREPARATION OF MONITOR BOTTLE
Before commencing operation, fill the monitor bottles (upper and lower) 200 mL of 10 % glycerol fluid or
water by following the procedure below.
1.
Remove the top and bottom drawers.
2.
Remove the cover on the monitor bottle.
The cover is fixed by clicking the top and bottom into
place. It can be removed by pulling the upper side
forward as shown in Fig. 1.
3.
Take the sensor for temperature display out of the
monitor bottle.
The sensor for temperature display is not fixed, and so
can be removed together with the cover.
Take out the recorder sensor from the upper monitor
bottle as well.
4.
Remove the fixing screws (2 screws) and take out the
monitor bottle.
5.
Remove the (screw-type) cap from the monitor bottle as
shown in Fig. 2.
6.
Pour 200 mL of 10 % glycerol fluid or water (using the
marking on the bottle) into the monitor bottle.
The displayed chamber temperature cannot be accurate
if the fluid volume in the monitor bottle is low. Check the
fluid volume regularly.
7.
Put the cap back on the monitor bottle and fix the
monitor bottle to its original location using the fixing
screws.
8.
Attach the sensor for temperature display to the monitor bottle and replace the cover.
Attach the recorder sensor to the upper monitor bottle as well.
9.
Replace the top and bottom drawers.
Monitor bottle
Monitor bottle
Cover
Fixing screw
Recorder sensor
(upper monitor bottle)
Sensor for
temp. display
Fig. 1
Loosen
Cap
Fig. 2