   DISPLAY OF CHAMBER TEMPERATURE 

 
The chamber temperature is set to 4 

o

C±1.5 

o

C at the factory.  The temperature on the temperature 

display shows the temperature of the sensor for temperature display attached to the monitor bottle (upper 
and lower). 
By pressing the temperature display changeover key, the displayed temperature is changed in the 
following sequence: upper temperature, lower temperature, average of upper and lower temperature.  
Table 1 shows the procedure for changeover. 

<Important> 

The upper temperature represents the temperature of the upper monitor bottle and the lower temperature 
represents the temperature of the lower monitor bottle.  Accordingly, the displayed temperature is not 
always 4 

o

C±1.5 

o

C.    This is not a malfunction. 

 

Table 1.  Changeover of temperature display             : Lamp ON

 

    : Lamp OFF

 

 

Operation 

Key operated

Indication after operation 

Display mode

Plug-in the power supply 
plug and turn on the 
power switch. 

----- 

Average temperature
is displayed. 

 

 

UPPER 

 

 

LOWER 

Average 
temperature 

Press the temperature 
display changeover key. 

 
 
 

Upper temperature is
displayed. 

 

 

UPPER 

 

 

LOWER 

Upper 
temperature 

Press the temperature 
display changeover key. 

 
 
 

Lower temperature is
displayed. 

 

 

UPPER 

 LOWER 

Lower 
temperature 

Press the temperature 
display changeover key. 

 
 
 

Average temperature
is displayed. 

 

 

UPPER 

 

 

LOWER 

Average 
temperature 

5  Repeat from step 2. 

 

 

 

 
 

   DEFROSTING OF EVAPORATOR 

 
The following two defrosting methods are adopted in this unit.    Both of them are controlled automatically. 
 

 Cycle defrost 

To keep the chamber temperature stable, the compressor is turned on and off in cycles.    During the “off” 
period any frost which has accumulated on the evaporator is melted by a defrosting heater.    This will not 
have any significant effect on the chamber temperature. 
 

 Forced defrost 

When the ambient humidity is high, or when a large number of damp items are being stored in the 
chamber, there is a possibility that the cycle defrost may not be sufficient to remove all of the frost on the 
evaporator.    In this case, a forced defrost can be initiated. 
When the unit is operating under the forced defrost, the current chamber temperature and “dF” are 
displayed alternately on the temperature display. 
Once the forced defrost is completed, the normal operation will resume.    The temperature of blood bags 
(400 mL) in the chamber will rise by approximately 1 

o

C to 2 

o

C during the forced defrost. 

UPPER

LOWER

UPPER

LOWER

UPPER

LOWER

Содержание MBR-1405GR

Страница 1: ...05GR Please read the operating instructions carefully before using this product and keep the operating instructions for future use See page 35 for model number Operating Instructions Blood Bank Refrig...

Страница 2: ...RM TERMINAL 20 TEMPERATURE RECORDER Recorder unit 21 Loading the pen cartridge 22 Setting the start time 23 Replacing the recording chart 23 ALARM FUNCTIONS 24 SELF DIAGNOSTIC FUNCTIONS 25 ROUTINE MAI...

Страница 3: ...ructions is lost or the page order is incorrect or if the instructions are unclear or inaccurate No part of the operating instructions may be reproduced in any form without the express written permiss...

Страница 4: ...are illustrated in the following way WARNING Warning indicates a potentially hazardous situation which if not avoided could result in serious injury or death CAUTION Failure to observe CAUTION signs...

Страница 5: ...in a location where corrosive gases such as acids are present Installing the unit in a location where corrosive substances are present may cause electric components to corrode leading to leakage and...

Страница 6: ...ck or short circuit if the liquid is spilled Never damage the power supply cord or power supply plug by breaking adapting placing near a source of heat bending with force twisting pulling adding weigh...

Страница 7: ...arry out disassembly and disposal of the unit Leaving the unit in a location that can be accessed by third parties may result in unexpected accidents e g the unit may be used for unintended purposes D...

Страница 8: ...pe is hot Touching the evaporating pipe during or immediately after operation of the unit may cause injury Always store items in the drawers Storing items outside the drawers may cause the items to de...

Страница 9: ...ed engineer or service personnel should be allowed to open these covers This symbol indicates that caution is required Refer to product documentation for details This symbol indicates an earth This sy...

Страница 10: ...10 REFRIGERATOR COMPONENTS Refrigerator unit Back side 11 1 Fluorescent lamp page 27 2 3 4 6 7 8 9 10 Light switch 5 Front cover page 13 Temperature recorder Outer door Door gasket...

Страница 11: ...temperature or condensation around the access port 6 Levelling foot front 2 locations These are screw bolts used to install and fix the unit Adjust the height of the levelling feet by turning the scre...

Страница 12: ...re displayed alternately when the self diagnostic function detects any abnormality page 25 6 Up arrow key This key is used during calibration of the sensor for temperature display This key is for serv...

Страница 13: ...for details 3 Power switch This is a power switch of the unit I ON O OFF 4 Glow starter This is a glow starter for the fluorescent lamp It is recommended that the glow starter is replaced at the same...

Страница 14: ...the ambient temperature is subject to sudden changes it will not be possible to achieve a stable cooling performance A flat surface where the floor is also capable of bearing the total combined weight...

Страница 15: ...the casters are raised 5 mm to 10 mm above the floor surface Fig 1 In addition rotate the levelling feet slightly clockwise or anticlockwise and adjust them so that the unit is completely level When...

Страница 16: ...Remove the fixing screws 2 screws and take out the monitor bottle 5 Remove the screw type cap from the monitor bottle as shown in Fig 2 6 Pour 200 mL of 10 glycerol fluid or water using the marking o...

Страница 17: ...ure is set to 4 o C 1 5 o C at the factory Check that the chamber temperature reaches 4 o C 1 5 o C on the temperature display Check that the chamber temperature falls to 4 o C 1 5 o C on the temperat...

Страница 18: ...red items near to the air exhaust vent The stored items may become frozen Important Never block the air intake or exhaust vent Moreover never disrupt the air circulation Always place stored items so t...

Страница 19: ...e 3 Press the temperature display changeover key Lower temperature is displayed UPPER LOWER Lower temperature 4 Press the temperature display changeover key Average temperature is displayed UPPER LOWE...

Страница 20: ...essing the alarm buzzer stop key BUZZER since the remote alarm terminal is operated in conjunction with the alarm buzzer However the alarm mode of the remote alarm terminal is not cancelled by pressin...

Страница 21: ...n the power ON and OFF 9 Pilot lamp A green light is seen through the recording chart when the power is ON This lamp will not turn ON when the temperature recorder is being powered by the back up batt...

Страница 22: ...cated by the arrows to open the head clamp at A and B 5 As shown in 2 2 in Fig 2 position the pen cartridge so that the guide pins fit into the guide holes on the pen arm 6 As shown in 2 3 in Fig 2 pr...

Страница 23: ...acing the recording chart 1 Turn off the power switch 2 Slightly raise the end of the pen lifter and remove from the pen lifter stopper Then rotate clockwise until the pen top is on the pen lifter as...

Страница 24: ...Door check indicator lights up Intermittent tone with 2 minutes delay The alarm mode of remote alarm terminal is cancelled by pressing the alarm buzzer stop key BUZZER since the remote alarm terminal...

Страница 25: ...emp are displayed alternately If the lower sensor for temp display is disconnected Alarm indicator blinks E17 and chamber temp are displayed alternately If the lower sensor for temp display is short c...

Страница 26: ...larm operation Check the alarm operation regularly every 6 months to ensure that the alarm is activated in case of power failure The alarm operation can be checked by following the procedure below 1 P...

Страница 27: ...e power supply plug Before turning off the power switch turn off the light switch if the light switch is on 2 Move the items stored in the left hand drawers to the right hand drawers Then remove all t...

Страница 28: ...tion at least once a year For temperature calibration contact our sales representative or agent REPLACEMENT OF BATTERY Replace the battery for the power failure alarm every 3 years when the message F...

Страница 29: ...the supply circuit is activated A breaker of 15 A or more is recommended The fuse on the supply circuit is blown The outer door has been kept open for a long time The outer door is open An error code...

Страница 30: ...l life check with your local solid officials option or proper disposal Label indication is obliged to comply with Japanese battery regulation Label indication is obliged to comply with Taiwanese batte...

Страница 31: ...the chemical involved German Entsorgung von Altger ten und Batterien Nur f r die Europ ische Union und L nder mit Recyclingsystemen Dieses Symbol auf den Produkten der Verpackung und oder den Begleitd...

Страница 32: ...r sur les piles pictogramme du bas Si ce pictogramme est combin avec un symbole chimique il r pond galement aux exigences pos es par la Directive relative au produit chimique concern Spanish Eliminaci...

Страница 33: ...Neste caso estar em conformidade com o estabelecido na Diretiva referente aos produtos qu micos em causa Italian Smaltimento di vecchie apparecchiature e batterie usate Solo per Unione Europea e Nazi...

Страница 34: ...l Onderstaand symbool Dit symbool kan in combinatie met een chemisch symbool gebruikt worden In dit geval volstaan de eisen die zijn vastgesteld in de richtlijnen van de desbetreffende chemische stof...

Страница 35: ...FL40SSECW Glow starter FG 4P Alarms High temperature alarm Low temperature alarm Power failure alarm Door alarm Battery For power failure alarm DC 6 V 1100 mAh automatic recharge Remote alarm contact...

Страница 36: ...Contamination of the unit Unit interior No contamination Yes No Decontaminated Yes No Contaminated Yes No Others 3 Instructions for safe repair maintenance disposal of the unit a The unit is safe to...

Страница 37: ......

Страница 38: ......

Страница 39: ......

Страница 40: ...rating Instructions Printed in Japan LDCL038900 0 S0418 0 2018 04 01 1 1 1 Sakada Oizumi machi Ora gun Gunma 370 0596 Japan EU countries only PHC Corporation 2018 Nijverheidsweg 120 4879 AZ Etten Leur...

Отзывы: