2.5 LIFE TIME EXPECTANTCY NOTE:
Operating at extreme temperatures can and will dramatically reduce life expectancy. The expected
lifetimes above were observed in a dry environment where both the electronics and pressure fittings
were subject to atmosphere. XS-HTI-4 is not rated for operation beyond 175°C and 10000psi; accuracy
and quality will not be guaranteed.
2.6 FREQUENCY RESPONSE NOTE:
An increase in pressure applied to the sensor will result in a decrease of the pressure oscillator
frequency (approximately 3 PSI/Hz). Transient temperature effects will influence the pressure oscillator
frequency. Therefore, the temperature oscillator is used in tandem. An increase in temperature of
the sensor will result in an increase of the temperature oscillator frequency (approximately 0.1°C/Hz).
Pressure readings during temperature stability will result in optimal
performance.
Page 8
Specification Sheet