background image

Page 2

Table of Contents

CONTENTS

SECTION

1

2

3

MECHANICAL INSTRUCTIONS

 

1.1 Process Fluid Fitting ..........................................................................................................................  

 

1.2  Consumables  .........................................................................................................................................    

 

1.3 Microcap Attatchment ....................................................................................................................

   

XTALX DDQS1 SPECIFICATION SHEET
 

2.1  Specification  Hightlights  ...............................................................................................................           

 

2.2  Mechanical  Specifications  ...........................................................................................................         

 

2.3  Electrical  Specifications  .................................................................................................................          

 

2.4 Additional Specifications ..............................................................................................................

 

2.5 Life Time Expectancy Note ..........................................................................................................

 

2.6 Frequency Response Note ..........................................................................................................

 

2.5 DDQS1-10000-150 Specification Drawing .........................................................................

MacOS SOFTWARE/COMMUNICATIONS
 

3.1 Pressure Measurment Walk Through ...................................................................................

 
 

PAGE

3

4-5

6

7
7
7
7
7

8
9

10-12

Содержание XtalX DDQS1

Страница 1: ...www phasesensors com XtalX DDQS1 Quartz Pressure Transducer Mechanical User Manual...

Страница 2: ...nstitute for Nanotechnology 11421 Saskatchewan Drive Edmonton AB T6G 2M9 TELEPHONE 780 264 2659 JANUARY 2022 COPYRIGHT PHASE ADVANCED SENSOR SYSTEMS CORP XtalX DDSQ1 Quartz Pressure Transducer Mechani...

Страница 3: ...DDQS1 SPECIFICATION SHEET 2 1 Specification Hightlights 2 2 Mechanical Specifications 2 3 Electrical Specifications 2 4 Additional Specifications 2 5 Life Time Expectancy Note 2 6 Frequency Response N...

Страница 4: ...Master PN 5472A1 2 Turn the 11 16 wrench counter clockwise to loosen 3 Avoid if possible scraping angled metal surface and PTFE O ring groove with metal parts or process fitting threads 4 Take care to...

Страница 5: ...ctronics Housing Back Up Ring DN 014 2 PEEK 2 4 E H O Ring DN 014 2 FFKM 5 Process Side O Ring M14 x 1mm 1 PTFE Note The O rings in the above table are not compatible with acetone 1 2 CONSUMABLES The...

Страница 6: ...Page 5 Mechanical Instructions The placement of the O ring is shown below 2 1 Figure 1 1 Placement of O rings 4 3 5...

Страница 7: ...ou do not touch the membrane as it is fragile 3 Engage the threads by hand while gently pushing the process end fitting 4 Use a 11 16 hex wrench on the microcap and a spanner wrench on the feedthrough...

Страница 8: ...ME 230 ms Page 7 Specification Sheet SECTION 2 XTALX DDQS1 SPECIFICATION SHEET MAX CURRENT CONSUMPTION 40 mA LOW POWER STATE CURRENT DRAW 1s SAMPLE 25 C 1 212 mA LOW POWER STATE CURRENT DRAW 1s SAMPLE...

Страница 9: ...e guaranteed 2 6 FREQUENCY RESPONSE NOTE An increase in pressure applied to the sensor will result in a decrease of the pressure oscillator frequency approximately 3 PSI Hz Transient temperature effec...

Страница 10: ...t a n d o f f s H i g h a c c u r a c y d i g i t a l f r e q u e n c y c o u n t i n g c i r c u i t b o a r d E l e c t r o n i c s h o u s i n g s e a l o p t i o n 1 d o u b l e o r i n g b a c k...

Страница 11: ...into the Black Female port on the FTDI Cable If there is no male header attached to the DDQS transducer follow the following Silk Screen to FTDI color connection Matrix when soldering and or connectin...

Страница 12: ...Ports COM LPT Communications Port COM1 USB Serial Port COM4 The list of connections ports where the USB Serial device COM4 is the newly inserted FTDI cable 3 Download Putty Realterm or your preferred...

Страница 13: ...to the sensor 10 Continuous Measurement mode will return a measurement every 100 ms 11 Enter in a number from 1 9 to enter into autonomous measurement mode where a sample is given once every number o...

Отзывы: