Revision: 3.0.3
System Lock
Lock the instrument, Once the system is locked,
modification to the system is prohibited,
parameter is readable. "Unlock" using your de
password. The password is composed of 6 nu
Current Loop Mode Select
0.
4-20mA
s
1.
0-20mA
s
2.
4-20mA vs.Vel.
s
3.
4-20mA vs.Energy
se
Other different current output characteristics a
according to his actual requirements.
In two graphs shown above, flow F
4mA
indica
indicates the value that the user entered in W
a positive or negative flow value as long as th
CL 4mA Output Value
Set the CL output value according to the flo
at 4mA. The flow unit
’
s options are the
those in Window m31. Once "4-20mA vs.
selected in Window M56, the unit should be
m/s.
Menu
4
7
Menu
5
5
Menu
5
6
D116 Series Ultrason
locked, any
ed, but the
designated
numbers.
set up the 4-20mA output to. be flow rate mode
set up output range to be 0-20mA mode
set up the 4-20mA output to. be velocity mode
set up current loop output 4-20mAcorresponding ene
ristics are displayed in below figures. The user can select one
ndicates the value that the user entered in Window M57;
Window M58. In the 4-20mA modes, F
4mA
and F
20mA
can
s the two values are not the same.
flow value
he same as
s.Vel." is
d be set as
Sy s t e m L o c k [
* * * * Un lo c k e d * *
CL Mo d e Se le c t [
0 . 4 -
2 0 mA
CL 4 mA Ou t p u t Va
0 m3
sonic Flowmeter
Page 30 of 53
ng energy
one of them
57; and flow F
20mA
can be selected as
[ 4 7
* * * *
t [ 5 5
mA
t Va lu e
m3 / h