Operating Instructions
Contador
Page 36 of 40
Revision 6
© 2019 – Pfeuffer GmbH
Problem
Cause
Remedy
The device does not
start after pressing the
button
.
The feed container is missing.
The drawer is missing.
The electronics may be defective.
Place the feed container on.
Place the drawer.
Checked by Pfeuffer GmbH.
The device displays:
»no sample«
Device stops counting
during the counting
process.
The device has stopped feeding,
because no product is detected for
a period of one minute.
Mains voltage interference
The electronics may be defective.
Fill in counting product. Remove any
blockages.
Have the mains voltage checked by
an
electrician
.
Checked by Pfeuffer GmbH.
The device does not
switch off automatically.
Mains voltage interference
The electronics may be defective.
Have the mains voltage checked by
an
electrician
.
Checked by Pfeuffer GmbH.
The result of the count
is wrong.
The speed is too high.
The slider on the feed container is
set incorrectly.
Incorrect manual data input
(speed, size)
Device does not count correctly.
The electronics may be defective or
dirty.
The bag or bottle was filled several
times in continuous operation.
Reduce the speed.
Adjust the slider according to the
grain size.
Correct the data.
Carry out a text by means of a
reference sample.
Checked by Pfeuffer GmbH.
Keep a void zone, i. e. leave a filling
position free.
The display flashes.
The display is still in the setting
mode.
Press the button
.
The device displays:
»preset wrong«
Preset = 0,
preset must be at least 1.
Preset > 99950, max. preset during
counting is 99950.
Counting differences:
The preset is bigger than the count.
Set a larger preset.
Set a smaller preset.
Change the preset or count
accordingly.