12
4.4. Pumping Off Dry Gases
No special measures are necessary. The gas ballast valve
may remain closed.
5. What To Do In Case Of Breakdown?
Problem
Possible Causes
Remedy
Pump doesn't attain final pressure
Pump dirty
Operate pump for a longer period with open
Operating fluid dirty
gasballast valve or change operating fluid
Leak in system
Repair leak
Unusual operating noises
Damaged pump stages
Request repair by Pfeiffer Vacuum Service
Damaged coupling piece
Request repair by Pfeiffer Vacuum Service
Oil leaking from the foot of the pump Radial shaft seal defective
Request repair by Pfeiffer Vacuum Service
Pump doesn't start
Ambient temperature < 12 °C
Warm up pump
Dirty pump stages
Request cleaning by Pfeiffer Vacuum Service
Damaged pump stages
Request repair by Pfeiffer Vacuum Service
Further repairs can only be carried out by trained specialists in accordance with Service Instructions.
Gasballast valve helium tight and corrosive gas models
4.5. Pumping Off Vapours
To prevent their condensation in the pump,
vapours should only be pumped off with the
pump at operating temperature and the gas
ballast valve 45 open.
Observe pump water vapour tolerance levels
(see Section 8.0 Technical Data). When the process has been
completed, allow the pump to continue running for approx. 30
minutes for operating medium cleaning purposes.
CAUTION