5. What To Do In The Case Of Breakdowns? 

Problem

Possible Cause

Remedy

Pump doesn't attain final pressure

• Pump dirty

• Clean the pump

• Operating fluid dirty

• Operate pump for a longer period with open gas 

ballast valve or change operating fluid

• Leak in system

• Repair leak

Unusual operating noises

• Silencer dirty

• Clean the silencer (see Service Instructions).

• Pump system damage

• Request repair by PFEIFFER Service

Pump doesn't start

• Mains power missing

• Check mains voltage and motor switch 

• Ambient temperature <12 °C

• Warm up pump

• Dirty pump stages

• Request cleaning by PFEIFFER Service

• Damaged pump system

• Request repair by PFEIFFER Service

6. Maintenance 

When carrying out servicing work, take the following
precautions:

Always ensure the pump cannot be switched
on when carrying out any work on the pump. If
necessary, remove the pump from the system to
carry out inspection work.

Only dismantle the pump as far as is necessary in order to
repair defects.

Dispose of used operating fluid in compliance with local
regulations.

When working with synthetic operating media, toxic
substances and substances contaminated with corrosive
gases, the relevant instructions governing their use must
be observed.

Only use benzine or similar agents for cleaning pump
parts. Do not use soluble cleaning agents.

6.1. Replacing The Operating Fluid

The rate of deterioration of the operating fluid depends on the
pump applications.

The level of deterioration of organic operating fluids (for
example P3) can be read off from the colour scale in
accordance with DIN 51578 on the supplementary sheet
PK 800 219 BN/B, on request.
– The colour scale enables precise colour determination.

Fill a specimen in a test tube or some similar vessel and

test by holding against the light.

Where the discolouration is dark yellow to red brown
(equivalent to 4 ... 5 on the scale) change operating fluid.

Where organic operating fluids, for example P3,

are involved, the operating fluid should be
changed at least once a year!

Replacing The Operating Fluid

Switch off the pump.

Unscrew operating fluid drain screw 4 and drain operating
fluid.

The operating fluid temperature can be as high

as 80 °C. During maintenance and repair work,
process related toxic gases and vapours can
escape from the operating fluid which may
become contaminated with harmful substances

(radioactive, chemical etc.). 
Disposal of used operating fluid is subject to the relevant
regulations. 

Screw in operating fluid drain screw 4.

Allow pump to run for about 10 seconds with open vacuum
flange.

Drain off remaining operating fluid.

In the case of serious contamination, the operating fluid
will have to be changed several times.

Screw in operating fluid drain screw 4 with O-ring 49.

Fill in fresh operating fluid and check level as per 
section 4.1.

Operating fluid can contain substances from

the medium pumped. Operating fluid must be
disposed of in accordance with the local
respective regulations.

Please request safety instruction data sheets  for operating
fluids from PFEIFFER or download from the INTRANET/
INTERNET.

Any other repairs may only be carried out by qualified personnel in conjunction with the service instructions.

Содержание DUO 2.5 A

Страница 1: ...Operating Instructions Rotary Vane Vacuum Pumps UNO 005 A UNO DUO 2 5 A UNO DUO 2 5 AC PK 800 152 BGB F 9911...

Страница 2: ...em Modifications reserved Danger of burns from touching hot parts 1 Safety Precautions Danger of personal injury 1 Safety Precautions 2 1 1 For Your Orientation 2 2 Understanding The Pump 3 2 1 Main F...

Страница 3: ...um connection and the pumping system The valve closes off the vacuum equipment without delay when the pump is at a standstill and vents the pump The Rotary Vane Vacuum Pumps are in two versions availa...

Страница 4: ...uctance value of the accessory there is a reduction in volume flow rate For the intake lines metal or PVC hoses with small flanges are available for laboratory work 3 4 Connecting The Exhaust Side Pre...

Страница 5: ...values The single phase drive motors of the UNO 005 A 2 5 A and the DUO 2 5 A AC are fitted with a built in thermal protection switch A cable with connecting plug is fitted to the motor for mains con...

Страница 6: ...id is possible when the pump is in final vacuum operations If there is a noticeable loss of pump fluid the casing and the radial shaft seals should be checked for leaks see Service Instructions 4 2 Sw...

Страница 7: ...cer 40 is a nozzle inside the pump housing and cannot be altered when dirty it should either be cleaned or replaced see Service Instructions 7 1 7 2 7 4 Gas ballast valve 7 for Corrosive Gas Processes...

Страница 8: ...P3 can be read off from the colour scale in accordance with DIN 51578 on the supplementary sheet PK 800 219 BN B on request The colour scale enables precise colour determination Fill a specimen in a...

Страница 9: ...erials in accordance with the Code of Federal Regulations CFR 49 Part 173 240 Definition and Preparation We will carry out the decontamination and invoice this work to you if you have not attached thi...

Страница 10: ...mbar 5 10 2 5 10 2 5 0 10 3 Total with gas ballast mbar 1 1 6 0 10 3 Max pressure in continuous operation bar 1 5 1 5 1 5 Water vapour compatibility mbar 20 15 15 Water vapour capacity at 50 60 Hz g...

Страница 11: ...s catalogue contains details of a comprehensive accessory programme which can both extend the range of application of the pump and provide protection against wear and tear under extreme operating cond...

Страница 12: ...full part number When ordering spare parts please state additionally the unit type and unit number see rating plate Please use this list as an order form by taking a copy Spare parts UNO 005 A Standar...

Страница 13: ...he repair and or service of vacuum components will only be carried out if a correctly completed declaration has been submitted Non completion will result in delay The manufacturer could refuse to acce...

Страница 14: ...tate in basso le direttive aplicate le norme standardizzate nonch le norme e le specifiche nazionali utilizatte E E E ES S S S Declaraci n de conformidad al tenor de las siguientes Directivas de la UE...

Страница 15: ...G G GR R R R Produkt Product Produit Prodotto Producto Produkt Produkt Produto UNO 005 A UNO DUO 2 5 A UNO DUO 2 5 AC Angewendete Richtlinien harmonisierte Normen und angewendete nationale Normen in...

Страница 16: ...2 259 6243 U S A Pfeiffer Vacuum Inc 24 Trafalgar Square Nashua NH 03063 1988 USA telephone 1 603 578 6500 telefax 1 603 578 6550 Venezuela Secotec S A Apartado 3452 Caracas 1010 A telephone 58 2 573...

Отзывы: