2 Installation
The user and/or integrator is ultimately responsible for the pump installation and must apply the specific
safety instructions, in accordance with local regulations. Only authorized technicians trained in the rele-
vant safety regulations for ionizing environment can work on the pump.
► Check the compatibility of the cabinet with the pump model:
– ACP 28R pump with cabinet for ACP 28R pump (label on the cabinet).
– ACP 40R pump with cabinet for ACP 40R pump (label on the cabinet).
Installation
The pump needs to operate in a horizontal position resting on its feet, the pumping axis must be verti-
cal, and the pump inlet must be on top.
1. Determine where the pump will be placed.
2. Handle the pump using a hoist: chapter “Handling” of the ACP 28-40 operating instructions
3. Install the cabinet in a clean environment. The I/O main switch must be accessible by the opera-
tor.
80
m
m
m
ini
80 mm
mini
80
m
m
m
ini
10
0
mm
mi
ni
10
0
mm
mi
ni
80
m
m
mi
ni
Ventilation
To guarantee the characteristics and performances of the pump and the cabinet within the boundaries
of the operating conditions:
► Do not obstruct the air vents.
► Keep the pump and the cabinet away from fixed walls, at a minimum using the value indicated on
the diagram below.
► Ensure the air circulation around the pump and the cabinet.
Installation
5/12