Operation Manual AM 5200 (-A) E5200EN3 v 1.1 May 2013
29 (36)
7.5
Error messages
Error message
Solution
Ambient Temperature out of range
The ambient temperature is either too high or too
low for the AM 5200 (-A) to work properly.
Clean sensor lenses, see instructions
The sample sensors may be dirty. Please see
instructions in Section 8.1 below. Also, make sure
that the instrument is not placed in direct sunlight.
Emergency stop, see instructions
A mechanical failure was detected. Check for
obstructions such as straws or other objects. Switch
power off and on again. Consult technical support if
the message persists.
Measuring cell not empty, see
instructions
Some sample or foreign material remained in the
test cell after the lid opened to allow the sample to
flow out. For example, this may be caused by a very
wet sample. Inspect the bottom of the cell (with
catch drawer removed) to look for residue around
the cell bottom and RF lid. If this does not resolve
the issue, it may be necessary to clean the cell as
described in Section 8.2.
Moisture out of range
The sample moisture content is outside the limits
the AM 5200 (-A) was calibrated for, and results
may not be accurate.