Operation Manual AM 5200 (-A) E5200EN3 v 1.1 May 2013
23 (36)
7.4.3.1
Product Settings
In “Product Settings”, the user can change settings for existing products. There are restrictions on
editing officially approved products. These products will display a padlock symbol on the product
button.
Setting
Description
Name
The name displayed in the Products List of the analysis start screen, and the
results database.
Profile
The profile contains the specific calibration used for the product.
Sample ID
Choose whether to use Default setting (as set in “Analysis options” under
“General settings”) or a specific setting for this product.
Abbreviation
Shows the official abbreviation for the product.
Moisture Bias
Bias is the offset used to adjust to the reference method.
Moisture Slope
Slope is the skew used to adjust to the reference method.
Moisture
Locked
Can be toggled to allow/not allow bias/slope changes for moisture.
TW Active
“Yes” means that Specific Weight (TW/HLW) is calculated when analyzing
this product. “No” means it is not. To set it to “Yes”, it is necessary to adjust
bias. Some products are bias adjusted from factory. These are the only products
with Specific Weight activated from factory.
TW Bias
Bias is the offset used to adjust to the reference method.
TW Slope
Slope is the skew used to adjust to the reference method.
TW Locked
Can be toggled to allow/not allow bias/slope changes for test weight.
Auto Print
Choose the Default setting (as set in “Analysis options” under “General
settings”) or a specific setting for this product.