Operation Manual AM 5200 (-A) E5200EN3 v 1.1 May 2013
21 (36)
7.4.1.4
Maintenance
In “Maintenance” the user can access the available maintenance tasks. These features should only
be accessed in the event of technical difficulties and as directed by technical support. Detailed user
maintenance procedures are located in section 8.
Setting
Description
Check scale
This feature is to check and adjust the built-in balance used for moisture and
Test Weight/Hectoliter Weight determination. Please contact technical
support for more information and assistance.
Grain Weight
Check
Displays the weight factor, sample temperature and total weight for a sample.
Stability Check
Allows the user to access more technical information if required during
technical support. (Administrator only).
Requires power cycle.
System recovery
Select [Set a restore point] to save the system configurations and data to the SD
card. (Administrator only)
Control values
Shows the values of the various measurements for the empty cell.
Clean
Allows opening of funnel for access to the inside of the cell for cleaning.
Change Pass
Code
Allows the Administrator to change the Admin Pass Code. Enter the current
Admin Pass Code (Default is 142857) in the “Current Pass” field. Enter the
new desired pass code in the “New Pass” and “Confirmation field. If the
revised Admin Pass Code is lost, the user can call Perten for an access code to
reset the Admin Pass Code to the Default setting.
Test Serial Ports
Enables diagnostics of the serial ports; Printer, Perten, and Custom.