Owner´s Manual Permobil C500
R-net Diagnostics
86
3.11
Indicator Lamp Short
This occurs when the control system detects a short in either of the Indicator
Circuits.
7206 - Left Indicator Short.
720A - Right Indicator Short.
•
Check the indicators, cables and connections to the control system.
3.12 Indicator Lamp Failed
This occurs when the control system detects a failure in either of the Indicator
Circuits. This is most likely to be an indicator bulb failure.
7207 - Left Indicator Failed.
7208 - Right Indicator Failed.
•
Check the indicator bulbs, cables and connections to the control system.
3.16 DIME Error
This occurs when the control system detects an identification conflict between
two modules in the system.
If a new module has been introduced:
•
Disconnect the new module and cycle the power.
•
If no trip is present connect the new module to the system and cycle the
power.
•
If the trip reappears then the new module must be the cause of the problem.
If there has been no additions:
•
Disconnect one module at a time and cycle the power.
If the trip is still present after the above checks have been made, contact your
service agent.