Permobil C400 Corpus 3G Lowrider
Troubleshooting R-net
Latched Timeout
This occurs when the control system detects that the programmed block time has been exceeded. This can, for exam-
ple, be due to the signal units (joystick, main steering device, suction and blowing device, etc.) not having been used
frequently enough.
The error reference provides information on why the control system has left block mode.
• Cycle the voltage.
• Activate block mode.
If the error persists after the checks listed above, the signal unit may be defective. Read more in
Repairing defective
units
, Page 120.
Brake Lamp Short
This occurs when the control system detects a short circuit in the brake lamp electrical circuit. Read more about con-
nectors in section 2.3.
• Check the brake lamps, their cables and the connections to the control system.
Lamp Short
This occurs when the control system detects a short circuit in the electrical circuit of one of the lamps.
7205 - Short circuit left-hand lamp.
7209 - Short circuit right-hand lamp
• Check the lamps, their cables and the connections to the control system.
Indicator Lamp Short
This occurs when the control system detects a short circuit in the electrical circuit of one of the indicators.
7206 - Short circuit left indicator.
720A - Short circuit right indicator.
• Check the indicators, their cables and the connections to the control system.
Indicator Lamp Failed
This occurs when the control system detects an error in the electrical circuit of one of the indicators. This usually
means the indicator needs replacing.
7207 - Error in left indicator.
7208 - Error in right indicator.
• Check the indicators, their cables and the connections to the control system.
DIME Error
This occurs when the control system detects an ID conflict between two modules in the system.
If a new module has been added:
• Disconnect the new module and cycle the voltage.
• If no error occurs, connect the new module to the system and cycle the voltage.
• If the error recurs, the new module must be the cause of the problem.
If no new modules have been added:
• Disconnect one module at a time and cycle the voltage.
If the error persists after the checks listed above have been performed, consult your service contact or Permobil.
Memory Error
This is a non-specific memory error that may be caused by any of the system modules.
• Check all cables and connections.
• Cycle the voltage.
If the error persists and the system includes third-party modules:
• Disconnect all modules that do not come from PGDT and cycle the voltage.
If this has dealt with the error:
• Connect one third-party module at a time and cycle the voltage each time.
• If the error recurs after one of the voltage cycles, the last module to be connected must be defective.
117
Содержание C400
Страница 1: ...Service manual C400 Corpus 3G Lowrider ...
Страница 2: ......
Страница 4: ...Text ...
Страница 5: ......
Страница 8: ......
Страница 10: ...Permobil C400 Corpus 3G Lowrider Introduction 10 ...
Страница 12: ...Permobil C400 Corpus 3G Lowrider Serial Number labels 12 ...
Страница 94: ...Permobil C400 Corpus 3G Lowrider Repairs 94 ...
Страница 110: ...Permobil C400 Corpus 3G Lowrider Control system 110 ...
Страница 123: ...Permobil C400 Corpus 3G Lowrider Cabling Owerview CABLING OWERVIEW Chassis 124 Seat 126 123 ...
Страница 124: ...Permobil C400 Corpus 3G Lowrider Cabling Owerview Chassis 124 ...
Страница 125: ...Permobil C400 Corpus 3G Lowrider Cabling Owerview 125 ...
Страница 126: ...Permobil C400 Corpus 3G Lowrider Cabling Owerview Seat 126 ...
Страница 127: ...Permobil C400 Corpus 3G Lowrider Cabling Owerview 127 ...
Страница 128: ...Permobil C400 Corpus 3G Lowrider Cabling Owerview 128 ...
Страница 131: ......
Страница 132: ...A 205334 UK 0 www permobil com ...