58 . Raman IdentiCheck Getting Started Guide
Figure 16 System verification / calibration flow diagram
If wavelength calibration verification fails and the Description of the calibration report
spectrum states that peak wavelengths are outside limits, then wavelength and laser
frequency calibrations should be performed. When these calibrations have been completed
successfully, repeat the wavelength verification to ensure that the instrument has been
brought back within specification.
If the wavelength calibration verification fails because the spectral intensity is low, then this
may also be due to the calibration sample not being correctly positioned, optimally focused
or because the laser key is not switched on or the laser power is too low. Then repeat the
wavelength calibration verification to ensure that the instrument has been brought back
within specification. If the instrument fails the wavelength calibration verification again,
please contact PerkinElmer.
Wavelength Calibration
Verification
Laser Frequency
Calibration
Wavelength
Calibration
Finished.
Instrument ready
for use
Check Laser Key is
turned on and laser
power is set to
100%
Check position and
focus of polystyrene
sample
Completed
Fail
Peak
wavelengths
outside limits
Fail
Intensity
low
Pass
Содержание Raman IdentiCheck
Страница 1: ...Raman IdentiCheck Getting Started Guide MOLECULAR SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 Raman IdentiCheck Getting Started Guide ...
Страница 11: ...Safety Practices ...
Страница 26: ...26 Raman IdentiCheck Getting Started Guide ...
Страница 27: ...Raman History Theory and Instrumentation ...
Страница 34: ...34 Raman IdentiCheck Getting Started Guide ...
Страница 35: ...Introduction to the Raman IdentiCheck ...
Страница 50: ...50 Raman IdentiCheck Getting Started Guide ...
Страница 51: ...Using the Raman IdentiCheck ...
Страница 65: ...Maintenance and Troubleshooting ...
Страница 73: ...Appendix A Using your own PC ...
Страница 78: ...78 Raman IdentiCheck Getting Started Guide ...