Maintenance
177
Figure 5-14
Spray chamber, end cap, and associated parts. (Illustration shows
cross-flow nebulizer end cap assembly, Part No. N0770546.)
Cleaning the Scott-Type Spray Chamber
If you have been analyzing aqueous solutions, the spray chamber should be cleaned
in an acid bath. An ultrasonic bath may be used. Start with a solution of 5% nitric
acid or aqua regia and, if deposits persist, increase the acid concentration, up to 20%
Item
Description
1
Spray Chamber (Part No. N0770357)
2
O-Ring (Part No. 09902033)
3
End Cap Ring (Part No. 00473543)
4
End Cap (Part No. N0681688 not complete assembly)
5
Knurled Screw (2) (Part No. 00473539)
1
2
5
4
3