Installation
121
Setting the Torch Viewing Position
This procedure adjusts the torch viewing position for the highest signal intensity.
Perform this procedure when:
•
the instrument or software is first installed
•
the instrument is moved to a new location
•
the torch is removed or replaced
•
the plasma induction plates is replaced
1. Switch on the computer and spectrometer, and if required, the autosampler.
2. Switch on the exhaust ventilation system.
3. Start the ICP Winlab software.
4. In the
Options
menu, click on
Autosampler
, and make sure that the software is
configured for the autosampler tray you are using.
5. In the
Tools
menu, click on
Plasma Control
and ignite the plasma. Allow the
torch to warm up for half an hour before you start the alignment.
6. Prepare a solution containing 1 ppm manganese.
7. In the
Tools
menu, click on
Spectrometer Control
.
The Spectrometer Control window appears.
8. Select
Axial
or
Radial
. We recommend that you first do the alignment in one
viewing mode, then select the other viewing mode and repeat the procedure.
9. To view spectra collected during the procedure, open the
Spectra Display
window.
10. Click on
Align View
in the Spectrometer Control window to open the
Align
View
dialog.
11. Select the element and wavelength. This is the wavelength that the system will
use to measure the intensities at different viewing positions and locate the
position that gives the highest intensity. Your options are:
Содержание OPTIMA 8000
Страница 1: ...OPTIMA 8000 Customer Hardware and Service Guide ICP OPTICAL EMISSION Return to Document Menu...
Страница 2: ......
Страница 3: ...Optima 8000 Customer Hardware and Service Manual...
Страница 12: ...Contents 10...
Страница 18: ...16 Indicates the ON position of the main power switch Indicates alternating current...
Страница 24: ...22 Figure C Location of warning labels in the sample compartment 1 2...
Страница 28: ...26...
Страница 29: ...Safety Practices 1...
Страница 49: ...Preparing Your Laboratory 2...
Страница 61: ...Preparing Your Laboratory 60...
Страница 62: ...System Description 3...
Страница 78: ...System Description 78 Figure 3 8 GemCone nebulizer and end cap N0680343 also shown in cross sec tional view...
Страница 89: ...Installation 4...
Страница 136: ...Maintenance 5...
Страница 241: ...Troubleshooting 6...
Страница 264: ...Error Messages 7...
Страница 284: ......