Instrument Methods . 65
10.
Set the
Factor
by which the absorbance is multiplied to give the concentration using
the left and right arrows.
The available range is from 0.01 to 9999.
11.
Press
OK
to start measurements.
The Kinetics results screen is displayed.
12.
Insert the reference sample and press
.
This reference measurement will be used for all subsequent samples until repeated.
13.
Insert your first sample and then press
to start the run.
The running
Time
is displayed at the bottom of the screen, and absorbance data are
plotted on the graph as the measurement proceed.
The table below the graph displays
A
0
,
the absorbance value at t
0
(start of calculation);
A
n
,
the absorbance value at t
n
(finish of calculation);
dA
, the change in absorbance; the
Slope
and the regression parameter (
R
2
) of the calculated slope. The concentration is a
calculated using the parameters selected in Method Settings 2 (
dA
,
Final A
or
Slope
) –
the name of the
Result
table column depends on the units chosen in Step 9.
Use the left and right arrows to move the cursor and display the time and absorbance
value at measured data points.
Use the up and down arrows to zoom in or out.
Содержание Lambda XLS
Страница 1: ...LAMBDA XLS XLS User s Guide UV VIS SPECTROSCOPY...
Страница 4: ...Appendices 85 Appendix 1 Specifications 86 Appendix 2 Customer Care Contacts 87 Index 89...
Страница 5: ...Introduction...
Страница 8: ...8 Lambda XLS XLS User s Guide...
Страница 9: ...Safety Information...
Страница 14: ...14 Lambda XLS XLS User s Guide...
Страница 15: ...Installing your Instrument...
Страница 18: ...18 Lambda XLS XLS User s Guide...
Страница 19: ...Using your Instrument...
Страница 39: ...Instrument Methods...
Страница 71: ...Accessories...
Страница 80: ...80 Lambda XLS XLS User s Guide...
Страница 81: ...Maintenance...
Страница 85: ...Appendices...
Страница 88: ...88 Lambda XLS XLS User s Guide...