Instrument Methods . 47
5.
Select the
Mode
using the left and right arrows, and then press the down arrow.
The options are
Absorbance
or
%Transmission
.
6.
Press
OK
to start measurements.
The Spectrum results screen is displayed.
7.
Insert the Reference sample and press
.
This reference measurement will be used for all subsequent samples until repeated.
8.
Insert your first sample and press
.
A spectrum and a table of the Absorbance/%Transmission values at each peak are
displayed.
Use the left and right arrows to move the cursor along the graph. When it reaches a
peak, the peak height and width are displayed at the top of the screen. To zoom in on
the wavelength scale, use the up arrow key. (This automatically scales the
Absorbance/%Transmission axis, depending upon the Graph Scale option, and this is
retained for subsequent measurements.) To zoom out again, use the down arrow key.
9.
Repeat Step 8 for each sample.
Содержание Lambda XLS
Страница 1: ...LAMBDA XLS XLS User s Guide UV VIS SPECTROSCOPY...
Страница 4: ...Appendices 85 Appendix 1 Specifications 86 Appendix 2 Customer Care Contacts 87 Index 89...
Страница 5: ...Introduction...
Страница 8: ...8 Lambda XLS XLS User s Guide...
Страница 9: ...Safety Information...
Страница 14: ...14 Lambda XLS XLS User s Guide...
Страница 15: ...Installing your Instrument...
Страница 18: ...18 Lambda XLS XLS User s Guide...
Страница 19: ...Using your Instrument...
Страница 39: ...Instrument Methods...
Страница 71: ...Accessories...
Страница 80: ...80 Lambda XLS XLS User s Guide...
Страница 81: ...Maintenance...
Страница 85: ...Appendices...
Страница 88: ...88 Lambda XLS XLS User s Guide...