Instrument Methods . 61
17.
Insert your first sample and then press
.
The absorbance of the sample is measured, and its concentration is calculated and
displayed.
18.
Repeat Step 17 for each sample.
Press
to display the available method options, which you can select using the
keypad numbers:
1 Method Settings
Returns to the Method Settings screen.
2 Print
Prints the data on the results screen using the printer selected in
the Utilities folder.
3 Show Graph
Toggles the calibration graph on and off. The cursors display the
absorbance and corresponding concentration values of the last
measured sample.
7 Sample Number
Adds a prefix to the sample number and resets the incrementing
number to the desired value.
8 Save Method
Use the left and right arrows to select the folder you want to store
the method in (Favorites/Methods 1–9/SD Memory Card), then
press the down arrow and enter a name for the method. Press
Save
to save the method.
9 Auto-print
Toggles Auto-Print on and off.
Press
or
, or wait 20 seconds to exit the method options.
Press
to return to the Standard Methods folder after all your samples have been
measured.
Содержание Lambda XLS
Страница 1: ...LAMBDA XLS XLS User s Guide UV VIS SPECTROSCOPY...
Страница 4: ...Appendices 85 Appendix 1 Specifications 86 Appendix 2 Customer Care Contacts 87 Index 89...
Страница 5: ...Introduction...
Страница 8: ...8 Lambda XLS XLS User s Guide...
Страница 9: ...Safety Information...
Страница 14: ...14 Lambda XLS XLS User s Guide...
Страница 15: ...Installing your Instrument...
Страница 18: ...18 Lambda XLS XLS User s Guide...
Страница 19: ...Using your Instrument...
Страница 39: ...Instrument Methods...
Страница 71: ...Accessories...
Страница 80: ...80 Lambda XLS XLS User s Guide...
Страница 81: ...Maintenance...
Страница 85: ...Appendices...
Страница 88: ...88 Lambda XLS XLS User s Guide...