Flexar Analytical / FX-10 UHPLC Pumps
IQ/OQ Rev. D June 2009
1.2
Documentation
1.2.1
Equipment Manual
Part No.
Title
Flexar Analytical / FX-10 UHPLC Pump User’s Guide
1.2.2
Purchase Order Certification
The Flexar Analytical / FX-10 UHPLC Pump Template-SN
is in compliance with the specifications of
the purchase order.
Purchase Order Number
Template-PO
1.2.3
PerkinElmer Service Engineer Training
Persons authorized by PerkinElmer LAS to perform validations have been properly trained. Training is
documented, and a certificate is issued by PerkinElmer LAS to that effect.
See Attachment #1 for PerkinElmer LAS Training Certificate.
1.3
Installation Qualification
Verify the following list and mark as appropriate:
Description
Verified
1.3.1
The working environment is suitable with reference to the appropriate section in the
user’s manual
1.3.2
The correct basic equipment has been provided in accordance with the purchase
order
1.3.3
The packaging is undamaged
1.3.4
The instrument, PC and other supplied equipment are undamaged
1.3.5
Users Guide is identified
1.3.6
Manufacturer’s specifications included in manual
1.3.7
The power requirement settings of the instrument, PC etc., are compatible with those
of the working environment.
Note: It is the responsibility of Template-Company to
ensure proper voltage / current / frequency of the electrical line supplied for
operating the Flexar Analytical / FX-10 UHPLC Pump.
1.3.8
Adequate room for ventilation and servicing the equipment