Instruction Manual Thriller

®

 

PEQLAB_v0314_E 7 

CLEANING, MAINTENANCE AND REPAIR

 

Please make sure to disconnect the instrument from the main power if cleaning is desired. 
We recommend the use of a mild soap solution to clean the housing. If sample liquid has 
run into the boreholes of the blocks, they should be cleaned with a mild soap solution 
immediately and rinsed with distilled water. Mild alcohol-containing disinfectants may be 
used for decontamination. Maintenance of the Thriller

®

 is not required. 

In case a repair gets necessary the device must be sent to the manufacturer. 

The surface 

of the device should be decontaminated with a lint free cloth soaked with 70 % ethanol 
before sending.

 The drill holes of the thermoblock shall be cleaned with a 1.5 % 

hypochlorite solution with the help of a cotton bud. 

Please fill in and sign the 

decontamination certificate and send it to the manufacturer together with the device.

 

TECHNICAL DATA 

Temperature control range: 

 

4 °C (at 20 °C RT) to 99 °C in 1 °C steps 

Temperature accuracy: 

 

 0.5 °C 

 
Heating rate:  

 

approx. 5 °C/min 

Cooling rate: 

 

4 °C/min 99 °C to 20 °C 

 

 

1 °C/min 20 °C to 4 °C 

 
The heating and cooling rates may differ when using fluid-filled reaction tubes or at 
higher ambient temperature. 
 
Mix intensity in steps of 50 rpm: 
0.5 ml thermoblock 

 

300 to 1500 rpm 

1.5 and 2.0 ml thermoblock   

300 to 1400 rpm 

0.2 ml thermoblock 

 

300 to 1400 rpm 

Microplate thermoblock 

 

300 to 1400 rpm 

 
Mixing stroke: 

 

3 mm 

Programs: 

 

1 direct operation program (P0) 

 

 

9 programs with 9 steps each (P1 – P9) 

 
Max. total time: 

 

99 h 59 m or infinite 

Max. mixing activity/pause 
(interval function): 

 

99 m 59 s 

 
Thermoblocks: 

 

24 x 1.5 ml block 

 

 

24 x 0.5 ml block 

 

 

24 x 2.0 ml block 

 

 

96 x 0.2 ml block 

  

Microplate 

block 

Содержание Thriller

Страница 1: ...PEQLAB_V0709E Instruction Manual Bedienungsanleitung Thriller Thermoshaker Incubator Thermo Inkubationsmischer 0109D E Creating the future together v0314_E D Creating the future together...

Страница 2: ......

Страница 3: ...irect operation program P0 4 Programs P1 to P9 5 Use thermostat without mixing activities 5 Use thermostat with mixing activities 5 Mixing without incubating facility 6 Mixing at intervals 6 Short mix...

Страница 4: ...epair or replacement of the unit or refund of the purchase price at PEQLAB s discretion PEQLAB is not liable for any consequential damages PEQLAB reserves the right to alter the technical specificatio...

Страница 5: ...ler has a direct operation program P0 The parameters will be saved automatically after input If the setpoint temperature is changed the instrument will start heating or cooling immediately without hav...

Страница 6: ...matically Use the key to skip to the total time Use the or button to set the total time The Thriller could be programmed infinite when total time is set to the maximum The maximum set point is 99h 59m...

Страница 7: ...us change of mixing activity and mixing break If no interval is required set these parameters to OFF If an interval function is required use the or button to set the duration of the mixing activity Us...

Страница 8: ...Subsequently set the time of the mixing break e g 0 05m 5 seconds Start the program The Thriller will start operating at a change of mixing activity 0 10m and mixing break 0 05m in succession After th...

Страница 9: ...a 1 5 hypochlorite solution with the help of a cotton bud Please fill in and sign the decontamination certificate and send it to the manufacturer together with the device TECHNICAL DATA Temperature c...

Страница 10: ...Thriller Thermoshaker Incubator 91 7010 Thriller Thermoshaker Incubator 110 V version for USA 91 7010 US Thermoblock 24 x 0 5 ml micro test tubes 91 7110 Thermoblock 24 x 1 5 ml micro test tubes 91 7...

Страница 11: ...osed of with regular waste Please contact PEQLAB Biotechnologie GmbH or your local distributor for disposal of Thriller Thermoshaker Incubator Waste electronic equipment is not useless waste Environme...

Страница 12: ...Direktbedienprogramm P0 13 Speicherprogramme P1 bis P9 13 Temperieren ohne Mischaktivit t 14 Temperieren mit Mischaktivit t 14 Mischen ohne Temperieren 14 Intervallmischen 14 Short Mix 14 Speichern d...

Страница 13: ...Die PEQLAB GmbH verpflichtet sich zur Reparatur oder dem Ersatz des Ger tes bzw der R ckerstattung des Kaufpreises nach ihren Bedingungen PEQLAB haftet nicht f r Folgesch den die aus der Verwendung d...

Страница 14: ...n Parameter werden automatisch nach der Eingabe gespeichert Wird die Solltemperatur ver ndert beginnt das Ger t sofort mit dem Heizen oder K hlen ohne dass das Programm gestartet werden muss Sollen di...

Страница 15: ...gegeben werden Wird kein Intervall gew nscht die Einstellung auf OFF stellen Wird eine Intervallfunktion gew nscht zun chst mit der oder Taste die Dauer der Mischaktivit t einstellen Mit der Taste auf...

Страница 16: ...intensit t einstellen Programm starten Der Thriller temperiert bis zum Ablauf der Zeit Nach Ablauf der Zeit ert nt ein Signalton Mischen ohne Temperieren Temperatur auf die Einstellung OFF stellen Der...

Страница 17: ...blocks sollten mit Hilfe eines Wattest bchens mit einer 1 5 igen Hypochloritl sung gereinigt werden Dem Ger t ist beim Versand an den Hersteller die ausgef llte und unterschriebene Dekontaminationsbe...

Страница 18: ...llen Sie bitte das Dekontaminationszertifikat vollst ndig aus Lieferungen ohne dieses Formular m ssen wieder an Sie zur ck geschickt werden Alle R ckholungen m ssen im Vorfeld mit PEQLAB abgekl rt wer...

Страница 19: ...2 1900 A ENTSORGUNGSHINWEIS Elektrische Ger te die mit dem Symbol der durchgekreuzten M lltonne gekennzeichnet sind d rfen EU weit nicht mit dem normalen unsortierten Hausm ll entsorgt werden Bitte ko...

Страница 20: ...logie GmbH 6404 Polling Tel 43 0 5238 84 169 info peqlab at www peqlab at UK PEQLAB Ltd Southampton SO31 7ZN Freephone UK 0808 202 1302 info peqlab co uk www peqlab co uk USA PEQLAB LLC Wilmington DE...

Отзывы: