Instruction Manual Thriller
®
PEQLAB_v0314_E 4
OPERATING THE THRILLER
®
Switch on the Thriller
®
at the rear side of the instrument, the display will show the values
of the direct operation program P0 set last.
Set.
temp./Act. temp. Total
time
Mixing intensity
Mixing activity
Mixing break
Prog No./Prog step
60/22 01h00m
450
rpm
00:10m 00:05m
P0
Figure 2: Display of the Thriller
®
Direct operation program P0
After the instrument has been switched on the cursor will start flashing at the setpoint
temperature.
Use the + or – button to set the requested temperature. The Thriller
®
will start the
heating or cooling process automatically.
Use the
key to skip to the total time. Use the + or – button to set the total time. The
Thriller
®
could be programmed “infinite”, when total time is set to the maximum. The
maximum set point is 99h 59min, after this you can choose “inf” for infinite.
Use the
key to skip to the mixing intensity. Use the + or – button to set the mixing
intensity.
Use the
key to skip to the mixing time. Use the interval function to enter a
continuous change of mixing activity and mixing break. If no interval is required, set
these parameters to OFF. If an interval function is required, use the + or – button to set
the time of the mixing activity. Use the
button to skip to the mixing break, and then
use the + or – button to set the time of the break. If the mixing time has been set to
OFF, the mixing intensity set is used for continuous mixing.