3. Troubleshooting
3.3.9. “
” Error (Rotor Imbalance)
Tubes are not
inserted
symmetrically in
rotor holes?
No
Yes
Sample liquid in
tubes are not
properly
balanced?
No
Yes
Defective or
improperly
adjusted balance
sensor?
Yes
No
Rotor is stuck
when a run
starts?
Yes
No
20
Return to vendor for
repair.
Change Rotor or
Motor.
Adjust Balance sensor.
Make sure that the same
volume of liquid is in
each tube.
Load tubes
symmetrically.
Содержание PerfectSpin 24 R
Страница 1: ...Refrigerated Microcentrifuge PerfectSpin 24 R Service Manual Ver 1 0...
Страница 3: ...PerfectSpin 24 R Service Manual ii...
Страница 8: ...PerfectSpin 24 R Service Manual Figure 3 Wire Connection Drawing II 5...
Страница 10: ...PerfectSpin 24 R Service Manual Figure 4 Connector Locations of Power Board PCB 7...
Страница 26: ...PerfectSpin 24 R Service Manual 23 11 Lid cover is fully disassembled as shown above...
Страница 39: ...5 Spare Parts List 5 Spare Parts List Lid CoverAssembly 5 1 36...
Страница 41: ...5 Spare Parts List Top Case Assembly 5 2 38...
Страница 43: ...5 Spare Parts List Base Assembly 5 3 40...
Страница 46: ...43 PerfectSpin 24 R Service Manual 5 4 Solenoid Module Assembly...
Страница 48: ...PerfectSpin 24 R Service Manual Motor ModuleAssembly 5 5 45...
Страница 50: ...PerfectSpin 24 R Service Manual 5 6 Final Assembly 47...
Страница 52: ...PerfectSpin 24 R Service Manual Wire Drawing 5 7 49...
Страница 54: ...51 P N 402 lrmlab 10 Issue Date June 2009...