3. Troubleshooting
Error Messages
3.2.
12
Message
Cause
Action
“
” Error.
1.
Press
(START/STOP) knob
when Lid is open.
2.
Defective lid lock sensor.
1.
Close the lid.
2.
Change
Lid
lock
module.
3.
Return the unit for
service.
“
” Error.
(rotor
imbalance)
1.
Tubes
are
not
inserted
symmetrically in rotor holes.
2.
Sample liquid in tubes not properly
balanced.
3.
Defective or improperly adjusted
balance sensor.
4.
Rotor is stuck when a run starts.
1.
Load
tube
symmetrically.
2.
Make sure that the
same volume of liquid
is in each tube.
3.
Adjust balance sensor.
4.
Change Rotor or Motor.
5.
Return the unit for
service.
Содержание PerfectSpin 24 R
Страница 1: ...Refrigerated Microcentrifuge PerfectSpin 24 R Service Manual Ver 1 0...
Страница 3: ...PerfectSpin 24 R Service Manual ii...
Страница 8: ...PerfectSpin 24 R Service Manual Figure 3 Wire Connection Drawing II 5...
Страница 10: ...PerfectSpin 24 R Service Manual Figure 4 Connector Locations of Power Board PCB 7...
Страница 26: ...PerfectSpin 24 R Service Manual 23 11 Lid cover is fully disassembled as shown above...
Страница 39: ...5 Spare Parts List 5 Spare Parts List Lid CoverAssembly 5 1 36...
Страница 41: ...5 Spare Parts List Top Case Assembly 5 2 38...
Страница 43: ...5 Spare Parts List Base Assembly 5 3 40...
Страница 46: ...43 PerfectSpin 24 R Service Manual 5 4 Solenoid Module Assembly...
Страница 48: ...PerfectSpin 24 R Service Manual Motor ModuleAssembly 5 5 45...
Страница 50: ...PerfectSpin 24 R Service Manual 5 6 Final Assembly 47...
Страница 52: ...PerfectSpin 24 R Service Manual Wire Drawing 5 7 49...
Страница 54: ...51 P N 402 lrmlab 10 Issue Date June 2009...