3. Troubleshooting
3.3.4. Can’t reach temperature
Is the pre-set
temperature higher than
the room temperature
when the rotor is not
running?
Yes
No
No
Compressor can not start?
Yes
No
Compressor will start,
but can’t reach
temperature?
Yes
Inspect whether the
condenser fan or
compressor fan is
activated?
No
Change the Power board.
No
Yes
Yes
Inspect whether
the condenser is
dirty?
No
Yes
Inspect whether the air
inlet and air outlet are
blocked?
No
16
Return to vendor for repair.
Remove the blocking objects.
Clean the Condenser.
Return to vendor for repair.
Return to vendor for repair.
Return to vendor for repair.
Reset the pre-set temperature
or start the spinning of Rotor.
Содержание PerfectSpin 24 R
Страница 1: ...Refrigerated Microcentrifuge PerfectSpin 24 R Service Manual Ver 1 0...
Страница 3: ...PerfectSpin 24 R Service Manual ii...
Страница 8: ...PerfectSpin 24 R Service Manual Figure 3 Wire Connection Drawing II 5...
Страница 10: ...PerfectSpin 24 R Service Manual Figure 4 Connector Locations of Power Board PCB 7...
Страница 26: ...PerfectSpin 24 R Service Manual 23 11 Lid cover is fully disassembled as shown above...
Страница 39: ...5 Spare Parts List 5 Spare Parts List Lid CoverAssembly 5 1 36...
Страница 41: ...5 Spare Parts List Top Case Assembly 5 2 38...
Страница 43: ...5 Spare Parts List Base Assembly 5 3 40...
Страница 46: ...43 PerfectSpin 24 R Service Manual 5 4 Solenoid Module Assembly...
Страница 48: ...PerfectSpin 24 R Service Manual Motor ModuleAssembly 5 5 45...
Страница 50: ...PerfectSpin 24 R Service Manual 5 6 Final Assembly 47...
Страница 52: ...PerfectSpin 24 R Service Manual Wire Drawing 5 7 49...
Страница 54: ...51 P N 402 lrmlab 10 Issue Date June 2009...